Technical Information
93
Immunity Test IEC 60601 Level
Compliance Level
Electromagnetic
Environment
Guidance
Power
frequency
(50/60 Hz)
magnetic field,
IEC 61000-4-8
30 A/m
30 A/m
Power
frequency
magnetic fields
should be at
levels
characteristic of
a typical
location in a
typical
commercial or
hospital
environment.
NOTE:
U
T
is the AC mains voltage prior to application of the test level.
Table 7: Electromagnetic Immunity
Immunity Test
IEC 60601
Level
Compliance
Level
Electromagnetic Environment
Guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of ReDS
System (including cables) than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter (see below).