14
K900000000SE 022628A0 071013
2.3 Measurement and Accuracy Ranges
All data relative to concentration accuracies are referred to an instrument operating at a constant temperature
within the correct operating range (-5°C .. +45°C), being in operation for at least 15 minutes, powered by its
internal battery and after completion of auto-zero procedure.
Notes:
(1)
The maximum CO
2
value displayed depends on the type of fuel.
(2)
Stated precision includes error of the external sensor RTD Pt100 class A DIN 43760 (1980).
(3)
Stated precision includes error of the external sensor type K thermocouple class 1 IEC584.
(4)
Pressures greater than 750 hPa may permanently damage sensors or impair their characteristics.
1
2
3
4
5
6
7
8
MEASUREMENT
SENSOR
RANGE
RESOLUTION
ACCURACY
O
2
Electrochemical sensor
0 .. 25.0% vol
0.1% vol
±0.2% vol
CO
with H
2
compensation
Electrochemical sensor
0 .. 8000 ppm
1 ppm
±10 ppm 0 .. 200 ppm
±5% measured value 201 .. 2000 ppm
±10% measured value 2001 .. 8000 ppm
diluted
Electrochemical sensor
0.15 .. 5.00% vol
0.01% vol
±20% measured value
CO Low range
with H
2
compensation
Electrochemical sensor
0 .. 500 ppm
0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured value 40.1 .. 500.0 ppm
diluted
Electrochemical sensor
100 .. 3125 ppm
10 ppm
±20% measured value
CO
Mid range
Electrochemical sensor
0 .. 20000 ppm
1 ppm
±100 ppm 0 .. 2000 ppm
±5% measured value 2001 .. 4000 ppm
±10% measured value 4001 .. 20000 ppm
diluted
Electrochemical sensor
0.3 .. 12.5% vol
0.01% vol
±20% measured value
CO
Hi range
Electrochemical sensor
0 .. 10.00% vol
0.01% vol
±0.1% vol 0 .. 2.00 %
±5% measured value 2.01 .. 10.00 %
NO
Electrochemical sensor
0 .. 5000 ppm
1 ppm
±5 ppm 0 .. 100 ppm
±5% measured value 101 .. 5000 ppm
NO Low range
Electrochemical sensor
0 .. 500 ppm
0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured value 40.1 .. 500.0 ppm
NOx
Calculated
SO
2
Electrochemical sensor
0 .. 5000 ppm
1 ppm
±5 ppm 0 .. 100 ppm
±5% measured value 101 .. 5000 ppm
SO
2
Low range
Electrochemical sensor
0 .. 500 ppm
0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured value 40.1 .. 500.0 ppm
NO
2
Electrochemical sensor
0 .. 1000 ppm
1 ppm
±5 ppm 0 .. 100 ppm
±5% measured value 101 .. 1000 ppm
NO
2
Low range
Electrochemical sensor
0 .. 500 ppm
0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured value 40.1 .. 500.0 ppm
CxHy
Pellistor sensor
0 .. 5.00% vol
0.01% vol
±0.25% vol
CO
2
Calculated
0 .. 99.9% vol
0.1% vol
CO
2
NDIR sensor
0 .. 20.0% vol
0.1% vol
±0.3% vol 0.00 .. 6.00 %
±5% measured value 6.01 .. 20.0 %
Air temperature
Pt100 sensor
-20.0 .. 120.0 °C
0.1 °C
±0.5 °C
Flue gas temperature
TcK sensor
-100.0 .. 1250.0 °C
0.1 °C
±0.5 °C 0 .. 100 °C
±0.5% measured value 101 .. 1250 °C
Pressure
(draught & differential)
Piezoelectric sensor
-10.00 .. 200.00 hPa
0.01 hPa
±1% measured value -10.00 .. -2.01 hPa
±2 Pa
-200 .. 200 Pa
±1% measured value 2.01 .. 200.00 hPa
Differential temperature
Calculated
0 .. 1250.0 °C
0.1 °C
Air index
Calculated
0.00 .. 9.50
0.01
Stack loss
Calculated
0.0 .. 100.0 %
0.1 %
Efficiency
Calculated
0.0 .. 100.0 %
0.1 %
Efficiency
(condensing)
Calculated
0.0 .. 120.0 %
0.1 %
Smoke index
External instrument
0 .. 9
Excess air
Calculated
0 .. 850 %
1 %
CO
high immunity H
2
Electrochemical sensor
0 .. 8000 ppm
1 ppm
±20 ppm 0 .. 400 ppm
±5% measured value 401 .. 4000 ppm
±10% measured value 4001 .. 8000 ppm
(1)
(2)
(4)
(3)