2.0 TECHNICAL SPECIFICATIONS
2.1 Technical Specifications
Autozero:
Automatic autozero cycle (with probe inserted in the stack except for Chemist
400B and Chemist 401 versions).
Dilution:
Expansion system of the CO sensor measuring range up to 50,000ppm
(5.00%v/v) programmable as a simple protection of the CO sensor with triggering
threshold programmable by the user. Preset triggering threshold at 1,500 ppm.
Self-diagnosis:
All the functions and internal functions are checked and anomalies signalled.
Type of combustible:
7 predefined by the factory and 16 that can be programmed by the user.
Power:
Li-Ion battery pack with internal protection circuit.
Battery charger:
External battery charger.
Charging time:
2 hours for charging from 0% to 90% (3 hours for 100% charge).
Instrument working time:
18 hours of non-stop operation (excluding printing).
Printer:
Internal, 24-column impact printer, using ordinary paper (roll 18 m long and
57 mm wide).
Printer powered:
By the analyser batteries.
Printer autonomy:
Up to 40 analysis reports with the batteries fully charged.
Internal data memory:
300 complete data analyses, time and name of the customer can be stored.
User data:
3 programmable user names.
Print-out heading:
4 lines x 24 characters, customisable by the user.
Display:
Graphic backlit LCD, measuring 42 x 60 mm.
Communication port:
USB with mini-USB connector.
Bluetooth (optional):
Communication range: <100 meters (free field)
Manufacturer: FREE2MOVE
Model: F2M03GLA - CLASS 1
Bluetooth SIG Qualification Design (QDL) Certificate: B012541
Certification EC-R & TTE: 0681
Line filter:
With replaceable cartridge, 99% efficient with 20um particles.
Suction pump:
1.2 l/min heads at the flue up to 135mbar.
Condensate trap:
Outside the instrument.
Carbon black:
Using an external hand pump; it is possible to enter and print the Bacharach
index.
Leak test:
Gas pipes tested for leaks with separate printout of the result, by means of the
attachment AACKT02, according to UNI 7129 (new systems) and UNI 11137:
2012 (existing systems), with automatic calculation of pipe volume.
Condensing boiler efficiency: Automatic recognition of the condensing boiler, with calculation and printout of
efficiency (>100%) on the LHV (Lower Heating Value) in accordance with
UNI10389-1.
Environmental gases:
Measurement and separate printout of the ambient CO and NO values.
Draught test:
Draught tested as per the UNI 10845 standard. Using the external draught gauge
AACDP02 the resolution is 0.1 Pa with 0.5 Pa accuracy.
Operating temperature range: -5°C to +45°C
Storage temperature range: -20°C to +50°C
Operating humidity range: 20% to 80% RH
Protection grade:
IP42
Air pressure:
Atmospheric
Outer dimensions: Analyser: 30,7 x 10,5 x 9,6 cm (W x H x D)
Case:
48.2 x 37.5 x 16 cm (W x H x D)
Weight:
Analyser: ~ 1.1 kg
Compliant with the European standard EN50379-1 and EN50379-2 for the following measurements:
O
2
CO medium
NO
Temperature (flue gas)
Temperature (supply air)
Pressure (draft)
Pressure (differential)
1
2
3
4
5
6
7
8