71
K900000000SE 022628A0 071013
CHOOSE THE SENSOR TO BE RECALIBRATED AND PROCEED AS DESCRIBED (CO SENSOR
EXAMPLE):
•
Connect the known concentration gas cylinder to the instrument as shown in the following scheme:
▲
Operator
Report header setup
Micromanometer
Language
►Calibration
CONFIGURATION
PASSWORD
0 0 0 0
PASSWORD
1 1 1 1
►CO
NO
NO
2
CALIBRATION
CO
►NO
NO
2
CALIBRATION
CO
NO
►NO
2
CALIBRATION
Action calibrate
►Applied 100.0
P
Measured 0
P
Is 2.22 uA
Ia 0.17 uA
Status original
CO CALIBRATION
Action calibrate
►Applied 100.0
P
Measured 0
P
Is 0.21 uA
Status original
NO CALIBRATION
Action calibrate
►Applied 100.0
P
Measured 0
P
Is 1.19 uA
Status original
NO
2
CALIBRATION
Enter the recalibration menu password
1111.
Installed sensors which can be recalibrated are shown, and
can be chosen for recalibration.
In recalibration screenshot, information about the calibration
in use and sensor output are displayed.
Action:
selection of action to make
calibrate:
save new calibration
set original:
bring
back
factory
original
calibration
set user:
bring back last user calibration
done
Applied:
selection of cylinder gas concentration
Measured: Actual sensor reading
Is:
'Is' current from the sensor
Ia:
'Ia' current from the sensor
Status:
Shows calibration status:
original:
factory original calibration in use
user:
user calibration in use
saving:
busy saving calibration
user cal OK:
user calibration successful
cal error:
user calibration error
orig cal ok:
restore of original cal successful
GAS CYLINDER
COMBUSTION ANALYZER
P-
P+
A
FLOW METER
0,5 l/m
WARNING!
Adequate ventilation must be provided when working with toxic gases, particularly the flow
meter and instrument outputs must be evacuated by a ventilation system.