11.4 Vital signs measurement
General
Blood pressure measurement
Fault
Cause
Remedy
Vital signs tab is not
displayed
USB connecting cable of the
SmartBucket not connected to the
monitor
• Connect the USB connecting cable of the
SmartBucket to the monitor
• Restart the device
SmartBucket not activated
Administrator:
Activate SmartBucket:
menu
seca mVSA 535
is being
operated on a wheeled
stand: Measuring results
for vital signs are
implausible
• Original seca USB connecting cable
not being used
• USB extension cable also in use
• Use original seca USB connecting cable
(in scope of delivery for the wheeled stand)
• Do not use a USB extension cable
Fault
Cause
Remedy
Implausible measuring
results
Excessive patient movement
• Ask the patient to move as little as possi
-
ble
Incorrect blood pressure cuff used
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Blood pressure cuff not applied
correctly
Correctly apply blood pressure cuff, see
instructions for use for blood pressure cuff
Blood pressure cuff applied to an
extremity to which there is intravenous
access
Apply blood pressure cuff to a different
extremity
Insufficient cuff pressure
Incorrect blood pressure cuff used
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Blood pressure cuff or compressed air
tube leaking
Dispose of blood pressure cuff, use
replacement
Pump in device faulty
Do not continue using device and have it
repaired by seca Service
No option to select
Multiple measurement in
the NIBP settings dialog
window
Administrator has configured a
measuring profile. The name of the
measuring profile is displayed.
Not a malfunction
• Press the name of the measuring profile
•
Starting a multiple measurement
Blood pressure values do
not appear in the EMR
system or in the
seca analytics 115
PC
software
Multiple measurement was performed
For current measurement:
•
and
Administrator:
• Specify Individual measurement as preset