Hygiene treatment • 103
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
Temperature probe (red/blue) with cable
As required
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with a mild soap solution
3. Clean temperature probe
4. Shake out temperature probe so that no liquid remains
in it
5. Allow to air-dry for approx. 30 minutes
Probe holder (red/blue)
As required
1.
2. Moisten a cotton bud with a mild soap solution
3. Wipe over all the surfaces of the probe holder
In-ear thermometer
After every
use
In-ear thermometer and cable:
1. Do
not
discard probe cover
2. Moisten a soft cloth with a mild soap solution: Ratio of
water to soap solution: 20:1, temperature: Max. 55 °C
(130 °F)
3. Wring out the cloth so that no excess liquid can
penetrate the in-ear thermometer
4. Wipe over in-ear thermometer and cable
5. Dry the in-ear thermometer and cable using a lint-free
cloth
6. Discard probe cover
As required
Measuring head and lens:
1. Carefully remove all foreign particles using a cloth
moistened with alcohol (70 % isopropanol)
2. Wipe dry the lens at the tip of the measuring head with
a lint-free cloth (e.g. spectacle-cleaning cloth)
3. Make sure that the lens at the tip of the measuring
head does not have any fingerprints or marks on it
4. Allow the in-ear thermometer to dry completely in air
Magazine holder for probe covers (in-ear
thermometer)
As required
2. Moisten a soft cloth or cotton bud with a mild soap
solution
3. Wipe over the surfaces of the magazine holder
4. Allow the magazine holder to dry completely in air
SpO
2
sensor with cable
Follow the manufacturer’s instructions for use
Patient cable for SpO
2
sensor
Follow the manufacturer’s instructions for use
Component (depending on version)
Interval
Cleaning