154 •
seca mVSA 535
/
seca mBCA 525
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
3. Press the
SpO
2
button.
The presets are displayed (in this case: Masimo SET
®
pulse oximetry):
4. Press the desired sensitivity:
NOTE
The Maximum setting (Masimo SET
®
pulse oximetry) is not available
as a preset. Select this setting for each measurement directly
Measuring oxygen saturation (SpO
.
5. Press the
Confirm
button.
The dialog window closes.
The modified settings are adopted.
Presets for temperature
(COVIDIEN
TM
FILAC
TM
3000 only)
1. Press the
Settings
tab.
Masimo SET
®
SpO
2
module
Mode
Indication
Normal
• Normal perfusion
• Mild perfusion disorders
Adaptive Probe Off Detection
(APOD)
Vigorous patient movements
Maximum
• Poor perfusion
• Severely disrupted signal, for
example due to indoor lighting or
direct sunlight
seca SpO
2
module
Mode
Motion tolerance
Stable
High
Normal
Normal
Sensitive
Low