![Seca mBCA 525 Скачать руководство пользователя страница 189](http://html1.mh-extra.com/html/seca/mbca-525/mbca-525_instructions-for-use-manual_1230118189.webp)
Connecting to an EMR system • 189
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
6.4 Configuring data transfer to the EMR system
seca recommends performing the data transfer to the EMR system only in
collaboration with seca Service and the manufacturer of your EMR system.
NOTE
More information on this can be found in the
seca 103/452
system
instructions for use (from version 2.0 or higher).
6.5 Use of certificates
The device is prepared for the use of certificates. If you wish to use this
function, please contact seca Service.
6.6 Summary: Device characteristics when connected to EMR system
Function
Standalone mode (factory setting)/
connection to
seca analytics 115
Connection to EMR system via
seca connect 103
User identification (measure)
seca user account, role: Physician
EMR user ID
User identification (manage)
seca user account, role: Admin
Device-specific user account, role:
Admin
Login
Select from list
• Scan barcode
• Manual
Password
Manual
Manual
Patient identification
Select from list
• Scan barcode
• Enter manually
Source of patient data
• Device
•
seca analytics 115
EMR system
Create/edit patient data on
device
Possible
Not possible
Storage location for
measurements
• Device database
•
seca analytics 115
• Unsubmitted measurements list
• EMR system
Unsubmitted measurements
list
Not available
Available