3 temperature
measurement
Manual temperature input
Masimo SET
®
SpO
2
measurement
General technical data
COVIDIEN
TM
GENIUS
®
3 in-ear thermometer
Measuring method
Direct mode
Measuring range (in-ear)
33 °C - 42 °C (91.4 °F - 107.6 °F)
Measuring time
Less than 2 sec
Resolution
0.1 °C; 0.1 °F
Accuracy
• Ambient temperature: 16 °C - 33 °C (60,8 °F - 91,4 °F)
Target temperature: 33 °C - 42 °C (91,4 °F - 107,6 °F)
± 0.3 °C (± 0.5 °F)
EN 60601-1: Medical electrical device, type BF
Type of protection against ingress of liquids
IP22
Manual temperature input
Device versions
Versions without integrated
temperature measurement
Measuring methods
Oral, axillary, rectal,
in-ear, on the skin, contactless
Temperature range
32 °C – 44 °C (89.6 °F – 111.2 °F)
Masimo SET
®
SpO
2
module
a
b
c
d
e
f
g
Measurement
Measuring method
Functional oxygen saturation
Spectrophotometry (red/infrared)
LED wavelength:
Red
Infrared
Maximum light output
This information may be of particular interest to clinicians
660 nm
905 nm
15 mW
Measuring time:
• Device switched on, sensor not applied
• Device switched off, sensor applied
≤ 12 sec
≤ 8 sec
Measuring range:
• SpO
2
• Pulse rate
• Perfusion index
0 % – 100 %
25 min
-1
– 240 min
-1
0.02 % – 20 %
Measuring accuracy:
• Measuring range
• SpO
2
(no patient movement)
• SpO
2
(patient movement)
• Pulse rate (no patient movement)
• Pulse rate (patient movement)
70 % – 100 %
70 % – 100 % ± 2 digits
h
0 % – 69 % not specified
70 % – 100 % ± 3 digits
0 ‰ – 69 % not specified
25 min
-1
– 240 min
-1
± 3 digits
25 min
-1
– 240 min
-1
± 5 digits
Low perfusion performance
• Pulse amplitude
• Transmission
• Oxygen saturation (SpO
2
)
• Pulse rate
> 0.02 %
> 5 %
± 2 digits
± 3 digits
Resolution:
• Oxygen saturation (SpO
2
)
• Pulse rate
1 %
1 min
-1