Modifying presets
NOTE
Your settings apply only to the current measurement procedure. If you
stop the measurement procedure
, the presets defined by the administrator become active again.
1. Ensure that the blood pressure cuff is not applied.
2. Press the
NIBP
field.
The
NIBP settings
dialog window opens.
The presets are displayed.
3. Press the desired measurement method.
• Inflation, continue at step 5.
• Deflation, continue at step 4.
4. Adapt initial pressure if necessary.
5. Press the desired measuring sequence:
• Single measurement
• Multiple measurement or – if a measuring profile has been configured
for the multiple measurement – name of the measuring profile
NOTE
If you have questions about configuring the multiple measurement,
contact your administrator.
6. Press the
Confirm
button.
The dialog window closes.
Changed settings are adopted for the current measurement.
7. Start the blood pressure measurement as described in the section
Preparing the blood pressure measurement
.