4. Press the
button.
The
seca connect 103: Workflow
page is displayed.
5. Specify the settings for the scan sequence for user and patient ID:
►
Sequence for ID scan fixed: 1. User, 2. Patient
►
Sequence for ID scan flexible
6. Specify whether measurements are to be possible even if the connection
to the EMR system is interrupted (offline measurements):
►
Do not allow offline measurement
►
Allow offline measurements
NOTE
Offline measurements are temporarily saved in the
Unsubmitted
measurements
list. When the connection to the EMR system is
active again, the measurements can be called up, evaluated and sub
-
Using the Unsubmitted measurements
.
7. Under
User-ID
, specify whether a password is required for login:
►
Password must be entered (recommended)
►
No password required
8. If desired, enter an
Organization ID
to assign the device to a department
in your institution, for example.
9. Adopt the settings:
a) Press the
button
b) Press the
button
The settings are saved.