![Seca mBCA 525 Скачать руководство пользователя страница 113](http://html1.mh-extra.com/html/seca/mbca-525/mbca-525_instructions-for-use-manual_1230118113.webp)
Troubleshooting • 113
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
COVIDIEN
TM
FILAC
TM
3000
temperature measurement
COVIDIEN
TM
GENIUS
®
3 temperature
measurement
Fault
Cause
Remedy
No temperature
measurement possible
Temperature module of the
SmartBucket not activated
Administrator:
Activate temperature module:
System
\
SmartBucket
tab
Implausible measuring
results,
temperature measurement
unsuccessful
Special patient condition, such as
hypothermia
• Assess patient vital signs using alternative
means
• Switch from predictive measurement
mode to direct mode
Rectal measurement performed with
blue temperature probe
Only perform rectal measurement with red
temperature probe
Oral/axillary measurement performed
with red temperature probe
Only perform oral/axillary measurement
with blue temperature probe
Patient activity before oral temperature
measurement:
• Physical exertion
• Eating/drinking
• Brushing teeth
• Smoking
Only perform oral temperature
measurement about 20 minutes after any
of these activities
Set measuring position does not match
actual measuring position
• Select measuring position to suit the
probe used
• Set correct measuring position on device
No probe cover used
• Disinfect the temperature probe
• Use probe cover
Temperature probe faulty
Dispose of temperature probe, use
replacement
Temperature probe cannot
be pushed completely into
the probe holder
Probe cover not discarded
• Carefully withdraw temperature probe and
probe cover from the probe holder
• Discard probe cover
• Push the temperature probe into the
probe cover
Fault
Cause
Remedy
No temperature
measurement possible
Temperature module of the
SmartBucket not activated
Administrator:
Activate temperature module:
System
\
SmartBucket
tab
Patient temperature exceeds
measuring range of in-ear thermometer
Assess patient vital signs using alternative
means
Patient temperature undershoots
measuring range of in-ear thermometer