Troubleshooting • 115
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
Manual temperature input
SpO
2
measurement
Fault
Cause
Remedy
Manual temperature input
not possible
Measured values outside permitted
range
Observe permitted temperature range,
Technical data:
Device has integrated temperature
measurement (probe or in-ear
thermometer)
• Not a malfunction
• Use the measuring equipment for the
device
Temperature module of the
SmartBucket activated
Administrator:
Deactivate temperature module:
System
\
SmartBucket
tab
Manual temperature
entries do not appear in
the
seca analytics 115
PC
software
Manual temperature entries are not
synchronized with the
seca analytics 115
PC software
Not a malfunction
• Enter values in the
seca analytics 115
PC
software manually
Fault
Cause
Remedy
Implausible measuring
results
Intra-aortic balloon pump affecting
pulse rate
Check pulse rate using ECG
Sensor damp
• Dry sensor
• Use dry sensor
Sensor not applied correctly
Apply sensor correctly, see instructions for
use for sensor
Strong ambient light
Cover application site with opaque material
Electromagnetic interference
• Switch off devices in proximity, isolate
interfering device
• Align interfering device differently or set up
in a different location
• Increase the distance between this device
and the interfering device
Poor perfusion
• Assess patient vital signs using alternative
means
• Apply blood pressure cuff to location with
better perfusion
Measurement not possible
Masimo SET
®
only:
Sensor service life expired
Use new Masimo SET
®
sensor.
Sensor or patient cable defective
Dispose of sensor or patient cable, use
spare part
Measurement not possible
Sensor or patient cable defective
Dispose of sensor or patient cable, use
spare part
Pulse is not found or is lost
Sensor too rigid
• Use suitable size of sensor
• Apply sensor to a different finger
Strong ambient light
Cover application site with opaque material
Poor perfusion
• Assess patient vital signs using alternative
means
• Apply sensor to a location with better
perfusion