5-39
Input type Pulse
A flowmeter with a digital pulse output is connected to a channel of the logger. The
flowmeter transmits one pulse per 16 litre.
input type: Pulse
Unit:
l
The flow can be measured per hour or per second. If per hour is selected, the
logging interval must be higher than 15 minutes. Otherwise, an error message
appears.
Determining the leading
sign for another channel
A flowmeter is connected to channel 2. In addition to the flow rate itself, the flow
direction shall be determined. For this purpose, a sensor which indicates the flow
direction by means of voltage values is connected to another channel of the logger.
Input type: Sign slave for other channel
First of all, it has to be specified which direction is indicated by the 5 V value. A
voltage value of 0 V automatically indicates the opposite direction. Furthermore, the
channel (flow), the determined flow direction applies to, has to be selected (in this
case Channel 2).