Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
Value Parameter:
value
Example
:
set a1ch1 2.03203332555e-007
The Calibration Coefficient a1 setting allows the user to enter the a1 calibration
coefficient for ADC channels 1 through 4 (
a1ch1
,
a1ch2
,
a1ch3
,
a1ch4
) as a
decimal value, usually in scientific notation. Up to 19 characters may be
entered.
IMPORTANT! Do not change the calibration coefficients unless you have
recalibrated the sensor or instructed to do so by Sea-Bird Scientific
personnel. Incorrect calibration coefficients will generate invalid data in
telemetry frames generating data using engineering units.
I
MMERSION
C
OEFFICIENT
Command Line Parameter:
imch[1-4]
Value Parameter:
value
Example
:
set imch1 1.368
The Immersion Coefficient setting allows the user to enter the immersion
calibration coefficient for ADC channels 1 through 4 (
a0ch1
,
a0ch2
,
a0ch3
,
a0ch4
) as a decimal value. Up to 6 characters may be entered.
IMPORTANT! Do not change the calibration coefficients unless you have
recalibrated the sensor or instructed to do so by Sea-Bird Scientific
personnel. Incorrect calibration coefficients will generate invalid data in
telemetry frames generating data using engineering units.