Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
D - OPERATION
Tab
1
Tab delimiter.
Im (
2
)
6 (max)
The Sea-Bird Scientific OPTIC2 fit type
Im
(immersion) coefficient for optical channel 2.
Tab
1
Tab delimiter.
Channel (
3
)
10
typical/
max
An AU formatted value representing the
sampled A/D counts from the third optical
channel.
Tab
1
Tab delimiter.
A0 (
3
)
14
(max)
The Sea-Bird Scientific OPTIC2 fit type
a0
coefficient for optical channel 3.
Tab
1
Tab delimiter.
A1 (
3
)
19
(max)
The Sea-Bird Scientific OPTIC2 fit type
a1
coefficient for optical channel 3.
Tab
1
Tab delimiter.
Im (
3
)
6 (max)
The Sea-Bird Scientific OPTIC2 fit type
Im
(immersion) coefficient for optical channel 3.
Tab
1
Tab delimiter.
Channel (
4
)
10
typical/
max
An AU formatted value representing the
sampled A/D counts from the fourth optical
channel.
A0 (
4
)
14
(max)
The Sea-Bird Scientific OPTIC2 fit type
a0
coefficient for optical channel 4.
Tab
1
Tab delimiter.
A1 (
4
)
19
(max)
The Sea-Bird Scientific OPTIC2 fit type
a1
coefficient for optical channel 4.
Tab
1
Tab delimiter.
Im (
4
)
6 (max)
The Sea-Bird Scientific OPTIC2 fit type
Im
(immersion) coefficient for optical channel 4.
TERMINATOR
2
This field indicates the end of the frame. The
frame is terminated by a carriage return/line
feed pair (0D
hex
and 0A
hex
).
Example Telemetry:
SATBI40001
2684550016
2147267103.1 2.03203332555e-007
1.368 2684315904
2147492578.4 1.95923384221e-007
1.410
2684407360
2147582763.7 2.03019013945e-007
1.365
2684127360
2147871011.6 1.97172313736e-007
1.354