![Sea-Bird Scientific OCR-504 Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/sea-bird-scientific/ocr-504/ocr-504_product-manual_1221405040.webp)
Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
P
OWER
C
OMMAND
The “
power
” command may be used to turn operational power on and off
during a command console session. Operational power supplies electronic
components in the instrument responsible for sensor data acquisition.
Powering down these components may reduce the instrument’s total power
consumption, if possible (some components cannot be powered down). The
instrument is otherwise fully operational, so communication is still possible.
When operational components are powered down, sensor data acquisition is
disabled.
This command requires one command line parameter. To turn on operational
power, use the “
power on
” command. To turn off operational power, use the
“
power off
” command. To query the operational power status, use the
“
power ?
” command. This will display a message similar to the one below:
Operational Power: on
IMPORTANT! Operational power will remain in the state set by this
command once the command console exits and normal operation
resumes.
S
ET
C
OMMAND
T
he “
set
” command modifies configuration parameters for the instrument.
These parameters affect various aspects of the instrument’s operation and can
be modified by the user to customize the instrument. For an OCR-504, if you
enter a “
set -?
” command, the following will be displayed:
Usage: set [parameter] [value]
set telbaud [telemetry baud rate (bps)]
set maxrate [maximum frame rate (Hz)]
set initsm [initialize silent mode (on|off)]
set initpd [initialize power down (on|off)]
set initat [initialize auto telemetry (on|off)]
set netmode [network mode (on|off)]
set netadd [network address (1-255)]
set netbaud [network baud rate (bps)]
set avg [on|off]
set usecal [on|off]
set immersed [on|off]
set latency [value]
set frametype [binary|short|long]
set a0ch[1-4] [value]
set a1ch[1-4] [value]
set imch[1-4] [value]
This command requires two command line parameters. The first parameter
specifies the configuration parameter to modify. The second specifies the new