Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
value to assign to the parameter. A list of all available configuration parameters
is shown above.
IMPORTANT! Be careful using this feature. Changes made to the OCR-
504 configuration parameters affect the way the instrument operates.
Before you modify any of configuration parameters, make sure you
understand the consequences of the change.
For more information on these parameters and their effect on your instrument’s
operation, see section
OCR-504 Configuration Parameters
below.
S
HOW
C
OMMAND
The “
show
” command displays configuration parameters for the instrument.
These parameters are modified by the “
set
” command explained above. If you
ente
r “
show -?
” at the command prompt, the following message will be
displayed:
Usage: show [parameter|all]
See help for the 'set' command for a list of
available parameters.
This command requires only one command line parameter, which is the same
as the fi
rst parameter of the “
set
” command. Using the “
show
” command in
this way displays the current value of the configuration parameter, even if it has
not yet been saved. You may also use “
all
” as the command line parameter
to show a complete list of all configuration parameters and their current values.
For example, using the “
show all
” command on your OCR-504 would display
something like this:
Sensor Type: irradiance
Frame Type: binary
Header: SATDI4
Telemetry Baud Rate: 57600 bps
Maximum Frame Rate: AUTO
Initialize Silent Mode: off
Initialize Power Down: off
Initialize Automatic Telemetry: on
Network Mode: off
Network Address: 100
Network Baud Rate: 76800 bps
Averaging: off
Immersed: on
Calibrated Output: on
Sensor Latency: 0
Cal coefficients:
Optical Channel 1:
a0: 2147582763.7 [2147582763]
a1: 2.03019013945e-007 [2.03019e-07]
a0: 1.365 [1.365]