![Sea-Bird Scientific OCR-504 Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/sea-bird-scientific/ocr-504/ocr-504_product-manual_1221405034.webp)
Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
D - OPERATION
Channel (
4
)
10
typical/
max
An AF formatted value representing the
calibrated output (engineering units) from the
fourth optical channel. For an irradiance
sensor, the units are “uW/cm
2
/nm”; for a
radiance sensor, the units are “uW/cm
2
/nm/
sr”.
TERMINATOR
2
This field indicates the end of the frame. The
frame is terminated by a carriage return/line
feed pair (0D
hex
and 0A
hex
).
Example Telemetry:
SATFI40001
5.6134 8.9193 14.6706
22.4710
ASCII
T
ELEMETRY
–
O
PTICAL
D
ATA IN
E
NGINEERING
U
NITS
W
ITH
C
ALIBRATION
C
OEFFICIENTS
Data is presented as tab-delimited ASCII text. Optical channel data is provided,
in engineering units, along with calibration coefficients. This is known as the
“long” frame type.
Relevant configuration parameters:
frametype: long
usecal: on
Field Name
Field
Size
(bytes)
Description
Instrument
6
A unique 6 character AS formatted string
denoting the start of a frame of telemetry. For
irradiance sensors, the instrument string is
“SATGI4”; for radiance sensors, the string is
“SATGR4”.
Serial Number
4
(1
– 10
allowed)
An AS/AI formatted string denoting the serial
number of the instrument. This field combined
with the INSTRUMENT field uniquely identifies
the instrument. This combination is known as
the frame header or synchronization string.
This is normally a four-character field.
Tab
1
Tab delimiter.
Channel (
1
)
10
typical/
max
An AF formatted value representing the
calibrated output (engineering units) from the
first optical channel. For an irradiance sensor,
the units
are “uW/cm
2
/nm”; for a radiance
sensor, the units are “uW/cm
2
/nm/sr”.
Tab
1
Tab delimiter.