Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
example, 0.5 Hz does not mean that half a frame is transmitted every second.
It means that every two seconds, a frame will begin transmitting.
I
NITIALIZE
S
ILENT
M
ODE
Command Line Parameter:
initsm
Normally, just after the instrument is powered up or reset, a start-up banner will
be output on the telemetry interface during the initialization sequence. The
messages in this banner, among other things, identify the instrument and
provide a copyright notification. If silent mode is enabled, this banner will not
be displayed. This ensures that no data will be transmitted on the telemetry
interface until, if running autonomously, normal operation begins and telemetry
output is available. Enabling silent mode does not mean that the telemetry
interface is disabled during initialization. The command console can still be
engaged. See section
D - OPERATION
for more information on the start-up
banner and the initialization sequence.
When modifying this paramet
er with the “
set
” command, you must enter either
“
on
” or “
off
” as the value parameter to enable or disable silent mode.
I
NITIALIZE
P
OWER
D
OWN
Command Line Parameter:
initpd
Near the end of the initialization sequence, operational power is normally
applied. However, the instrument can be configured to boot into a power
savings mode. With the initialize power down parameter enabled, operational
power will not be applied during initialization. This means that telemetry output
will be disabled when normal operation begins. See section
D - OPERATION
for more information on the initialization sequence.
When modifying this parameter with the “
set
” command, you must enter either
“
on
” or “
off
” as the value parameter to enable or disable the power savings
mode.
I
NITIALIZE
A
UTOMATIC
T
ELEMETRY
Command Line Parameter:
initat
For instruments running in autonomous operation only, the telemetry output
mode can be configured to start as free-running or polled once normal
operation begins. This means that once normal operation has begun and this
parameter is disabled (polled operation), telemetry output will not occur unless
the instrument is polled with the <Enter> or <Space> key commands.
Otherwise, telemetry output will be free-running in accordance with the
maximum frame rate configuration parameter. Of course, if operational power
is not applied, telemetry output is disabled altogether regardless of this
parameter. See
Autonomous Operation
in section
D - OPERATION
for more
information on telemetry output modes.
When modifying this parameter with the “
set
” command, you must enter either
“
on
” or “
off
” as the value parameter to enable or disable automatic telemetry.