Section 1 SeaFET V2 and SeapHOx V2 quick start guide
This quick start guide gives the steps necessary to make sure that the SeaFET V2 or
SeapHOx V2 sensor operates correctly and collects data before it is deployed.
•
SeaFET V2 measures pH only
•
SeapHOx V2 is a SeaFET with a SBE-37-SMP-ODO attached. It measures the
added parameters of conductivity, temperature, pressure, and dissolved oxygen. The
two sensors are integrated by the manufacturer.
The SeaFET V2 acts as a controller for the SBE-37, which sends the data it collects to
the SeaFET V2. Because the SBE-37 pump operates before each burst of data
collection, the manufacturer recommends that the shortest data collection interval is every
two minutes for the SeapHOx. The shortest data collection interval for a SeaFET is
6 seconds. Use the battery endurance calculator in the software to determine the length
of a deployment.
N O T I C E
Do not let the sensing elements get dry. If the sensor will not be deployed immediately, put the wet
cap in place and fill with clean seawater.
What's in the box:
SeaFET V2
•
Flow cells and fittings
•
Anti-fouling guard
•
CD—has software, calibration files, documentation
•
RS232 cable to connect to a PC
•
SeaFET V2 toolkit and O-rings
•
Hardware kit for boss seal
SeapHOx V2
•
Flow cells and fittings
•
Anti-fouling guard
•
CD—has software, calibration files, documentation
•
Y-cable to connect the SeaFET V2 and the SBE-37-
SMP-ODO
•
RS232 cable to connect the Y-cable to a PC
•
2.4 m cable to communicate directly with the SBE-37-
SMP-ODO only
•
SeaFET V2 toolkit and O-rings
•
Mounting brackets (2) to attach the SBE-37 to the
SeaFET V2
•
Plumbing kit and Triton
®
X-100 to clean MicroCAT flow
path.
•
Hardware kit for boss seal
The SeaFET V2 has 12 size D alkaline batteries installed. The 12 lithium batteries for the
SBE-37 are shipped separately.
1. SeapHOx V2
: Install the manufacturer-supplied lithium batteries in the SBE-37 (refer
to
on page 16 for details.)
a.
Remove the end flange of the SBE-37.
b.
Disconnect the battery holder and remove it from the sensor.
c.
Install new batteries.
d.
Connect the battery pack to the sensor again and install the end flange again.
2. SeapHOx V2
: Remove the yellow protective label from the SBE-37 intake and
exhaust ports.
3. SeapHOx V2
:
a.
Connect the SeaFET V2 and SBE-37 with the manufacturer-supplied Y-cable.
b.
Connect the Y-cable to the manufacturer-supplied data I/O cable.
4. SeaFET V2
:
•
Connect the sensor to the manufacturer-supplied data I/O cable.
3
Содержание Deep SeapHOx V2
Страница 2: ......
Страница 8: ...2 1 3 Illustrations Figure 1 SeaFET V2 dimensions Specifications 6 ...
Страница 9: ...Figure 2 Y cable for SeapHOx V2 Figure 3 Data I O cable Specifications 7 ...
Страница 12: ...Specifications 10 ...
Страница 16: ...Product overview 14 ...
Страница 26: ...Deployment and recovery 24 ...
Страница 42: ...Maintenance 40 ...
Страница 46: ...Reference 44 ...
Страница 50: ...Software reference 48 ...
Страница 55: ......