19.
Connect the sensor to the software and turn on the internal batteries.
The voltages in the sensor
Dashboard
for new batteries are 12V for the Main Battery
Voltage and 6V for the Isolated Circuit Voltage. If they are not, open the sensor and
correct the problem.
7.6 Remove or replace anti-fouling devices
C A U T I O N
AF24173 Anti-fouling devices contain bis(tributyltin) oxide. Wear rubber or latex gloves and eye
protection to replace these devices on the sensor. Wash hands with soap and water when
finished.
Read the precautions on the product label.
It is a violation of US federal law to use this product in a manner that is inconsistent with its label.
Anti-fouling devices are installed at the intake and exhaust ports of the pump. Remove
these as a first maintenance task to save the anti-fouling material for deployments.
1.
Remove the four Phillips-head screws that attach the conductivity cell guard to the
pressure housing.
2.
Carefully remove the guard.
Maintenance
35