Section 8 Reference
8.1 Terminal program setup and use
If necessary, use a terminal program to set up and operate the sensor.
1.
Use the test cable to connect the sensor to the PC and a 12V power supply.
2.
Start a terminal emulator program such as HyperTerminal
®
or Tera Term.
3.
Select "Serial" for the type of connection.
4.
Set up the connection at 19200 (default), 8 bits, no parity, 1 stop bit, flow control:
none.
8.2 Terminal commands
This is a reference for advanced users. The values of these commands are stored in the
sensor until the user changes them. Notes about terminal commands are listed below.
•
Commands are not case-sensitive. Use "Enter" to store a command.
•
The sensor sends an error message if a command is invalid.
•
The argument Y and 1 are both "Yes" and N and 0 are both "No." For example,
OutputSal=y and OutputSal=1 are equivalent.
•
If there is no communication with the sensor for 2 minutes, it goes into a low power
mode. Use "Enter" to start communication again.
•
Push the "Esc" key, then "Enter" to stop the sensor as it sends data.
•
If the user sent StartNow (autonomous mode) and the sensor is in operation or in
standby, the user can use the Status commands, TS, TPS, SL, QS, and Stop. For
example, if the user sends a DS to see status data, the sensor completes the current
measurement and then responds to the command. If OutputExecuted=Y, the sensor
will send "executing" messages until the measurement is complete.
•
If the user sent StartLater (autonomous mode) and the sensor is operation or in
standby, the user can use the Status commands, TS, TPS, SL, QS, and Stop. To
send other commands, enter the Stop command, then enter any other commands,
and send StartLater again.
Status
GetCD
show configuration
GetSD
show status
GetCC
show calibration coefficients
GetEC
show event counter
ResetEC
reset event counter
GetHD
show hardware
Help
shows list of available commands
DS
show status and configuration
DC
show calibration coefficients
General setup
DateTime=x
set clock.
Format is mmddyyyhhmmss
BaudRate=x
RS232 rates. Default is 19200. 600*, 1200*, 2400*, 4800, 9600, 19200, 38400, 57600, 115200.
*available only if no oxygen sensor is installed.
OutputExecutedTag=x
x=Y: show XML executing and executed tags. x=N: do not show
41
Содержание Deep SeapHOx V2
Страница 2: ......
Страница 8: ...2 1 3 Illustrations Figure 1 SeaFET V2 dimensions Specifications 6 ...
Страница 9: ...Figure 2 Y cable for SeapHOx V2 Figure 3 Data I O cable Specifications 7 ...
Страница 12: ...Specifications 10 ...
Страница 16: ...Product overview 14 ...
Страница 26: ...Deployment and recovery 24 ...
Страница 42: ...Maintenance 40 ...
Страница 46: ...Reference 44 ...
Страница 50: ...Software reference 48 ...
Страница 55: ......