5.5 Recover sensor from deployment
N O T I C E
Do not let the sensing elements get dry. If the sensor will not be deployed immediately, put the wet
cap in place and fill with clean seawater.
N O T I C E
Do not leave the sensor in direct sun. Heat over 35 °C can cause damage to the sensor.
N O T I C E
Do not let the potassium chloride (KCl) gel or the wet cap filling solution freeze. This will damage
the DuraFET and void the warranty.
When the sensor is removed from the water after a deployment:
1.
Use the software to turn off the sensor.
2.
Immediately remove the anti-fouling guard and attach the wet cap filled with clean
seawater, or, if deployed with the flow cells, make sure there is seawater in the wet
cap and plug the cells to keep fresh water away from the sensing elements.
3.
Disconnect the deployment Y-cable.
4.
Flush the bulkhead connector, end flanges, and pressure housing with fresh water.
5.
Completely dry the bulkhead connectors, then attach the dummy plugs.
6.
Attach the dummy plug and lock collar to the bulkhead connector.