10. Preparing the model

When using hard or hard-elastic material, the model is placed in the model cup 

4

 that has been filled with pellets to its upper sealing rim. When using 

soft-elastic material, the model is placed on the model platform 

6

. Ensure in both cases that the front teeth are facing to the right in the direction of 

the locking shaft 

3

 or to the left in the direction of the pressure chamber 

2

 in order to avoid a different material thickness between the quadrants. 

11. Programming using the scanner or by entering the code or heating time

A three-digit code is assigned to all our pressure moulding blanks, containing all required information on temperature, heating and cooling time. This 

code is printed in plaint text and as bar code on all our material except for material with a thickness of 0.1 mm or 0.15 mm. Once you have selected 

the desired material, press the button 

“code/scan”

. The scanner 

P

 is now activated. Now present the bar code of the selected material under the 

scanner until you hear an acoustic signal. You also have the option to enter the code manually: Enter the three digits and confirm by pressing “code/

scan” again. Another option is to enter the heating time manually. Doing so, press the key with the clock symbol 

Q

, enter the required heating time 

and confirm by pressing key 

Q

 again.

12. Heating

Once you have scanned, entered the code or the heating time, swivel the heater 

1

 over the fixed blank. The temperature will be automatically ad-

justed by a thermo element, which is visible by a change of the heater colour. 

13. Thermoforming

Acoustic signals indicate the end of the heating process (+5 sec to 0 sec = beeps / 0 sec to -5 sec = continuous tone). Once 0 sec are reached, the 
heater shall be swivelled backwards, upon reaching -5 sec, the heater is switched off automatically for safety reasons. Swivel the pressure chamber 

2

  

by a 180° turn to the right 

(pay attention to clean sealing surfaces)

 and turn the locking shaft 

C

 180° to the front. The pressure chamber is filled 

with air and the cooling time starts automatically which is indicated by a blue LED. 

14. Depressurizing

Optical signals (blue flashing LED) and acoustic signals (4 beeps) indicate the end of the cooling phase. Pressing the flashing button 

“air”

 depressur-

izes the pressure chamber, 4 further beeps after approx. 3 sec indicate the end of depressurizing. Turn now the locking shaft 

3

 180° backwards, 

push the locking ring 

  to the left and open the pressure chamber 

2

 

. Once the model has been removed, the device can be used for the 

next pressure moulding process.

15. Special programs for the use of the heating support

You have the option to program time intervals of 60/120/180/240 sec. For this, the pressure chamber shall be closed and the locking shaft shall be 

opened. These time intervals can be used for heating and bonding the occlusal surfaces using the heating support (REF 3452). Once you have entered 

the corresponding heating time, the heater with the attached heating support is swivelled forwards in the heating position. This operation can be car-

ried out several times depending on the material thickness.

16. Service/maintenance

The BIOSTAR

®

 device is generally maintenance-free. Ensure however to control the air tube with the filter element for any kind of debris or moisture. 

The filter element can be exchanged if necessary. The model cup 

4

 and the exhaust air holes as well as the model platform 

6

 can be cleaned by 

means of air nozzle or steam jet. Ensure the reflective surfaces in the pressure chamber are always clean for optimum heating results. The aluminium 

balancing ring 

L

 in the pressure chamber 

can be removed for cleaning by tipping. Doing so, clean and grease the sealing surfaces, the lateral 

sealing ring, the 4 rivets and springs. Make sure to refit the balancing ring on the springs and rivets with the sealing ring facing upwards. The whole 

unit shall be freely movable (see also Troubleshooting Measures). 

17. Service updates

The following applies to devices with serial numbers beginning with 70858 for the 230 V version, with 70527 for the 115 V version and with 70103 

for the 100 V version:
You have received a registration card with your device. Once you are registered, you will be automatically informed about updates that can be 

downloaded from our homepage 

https://www.scheu-dental.com/en/service/support-and-program/

. In order to access the SD card (see fig. 4) 

unscrew the cover on the left side of the device. Remove the SD card and erase any existing data or reformat your SD card. Extract the download file 

on the computer and copy the subdirectories “data” and “fw” to a blank and formatted SD card. Insert the SD card in the 

service interface

 

S

 on the 

left side of the device. Refit the cover of the service interface on the device. The update for the SD card is now complete.

The following applies to devices with serial numbers beginning with 70001-70857 for the 230 V version, with 70001-70526 for the 115 V version 

and 70001-70102 for the 100 V version: 
In these cases, you can use a USB flash drive for the updates: Ensure to use a blank and formatted USB flash drive. Extract the download file on the 

computer and copy the subdirectories “data” and “fw” to the USB flash drive. A video showing the procedure is available in the support area. Start 

the device press the 

“ok”

 button 5 times. You are asked to enter the password for the info menu. Enter the digit 1 and confirm by pressing 

“ok”

.  

By pressing the 

digit “1”

 again, the update starts. Once the update is complete, quit the service menu pressing the key 

“CE+home”

We reserve the right to make technical changes in the interest of progress. Reproduction, even in extracts, is subject to our written permission.

G

 

English

Содержание 3001

Страница 1: ...Bedienungsanleitung Operation Manual Mode d emploi Manual de Instrucciones Istruzioni per l uso S C H E U G R O U P ...

Страница 2: ...AC 230 V 7 8 115 V 100 V 7 R 50 60 Hz Leistung Power 750 VA B x H x T W x H x D 520 x 260 x 440 mm Gewicht Weight 14 kg Druck Pressure 9 max 6 bar 87 psi J J 9 R 9 7 R 8 7 T4A 230 V TT10A 100 V T10A 115 V Abb Fig 1 Abb Fig 2 230V Abb Fig 3 100 115V Abb Fig 4 S T 21 22 N ...

Страница 3: ...geschlossen werden Umwelt Das Gerät ist aus wiederverwendbaren oder recyclebaren Materialien hergestellt Gemäß der 2012 19 EU Elektro und Elektronik Altgeräte Richtlinie nimmt SCHEU DENTAL das Gerät bei freier Rücklieferung kostenlos zurück Alternativ kann es an einer Sammelstelle für das Recycling elektrischer und elektronischer Geräte abgegeben werden 2 Allgemeine Inbetriebnahme Zur optimalen Nu...

Страница 4: ...eckdose auch zwei Gerätesicherungen die die Elektronik vor Überspannungen schützen Zum evtl Austausch drücken Sie bitte den kleinen Clip unter der Netzsteckdose und entnehmen Sie das Sicherungsfach 8 Bei der Geräteausführung in 100 V 115 V befinden sich die Sicherungen in 2 separaten Halterungen R die Sie aufschrauben können Für 230 V Geräte verwenden Sie die Sicherung T4A 3100 940 für 115 V T10A ...

Страница 5: ...0 180 240 Sekunden programmiert werden Diese Zeiten dienen dem Erwärmen oder dem Verbund der Okklusalflächen unter Zuhilfenahme des Heizsupports REF 3452 Nach Programmierung der entsprechenden Heizzeit wird der Strahler mit angehängtem Heizsupport nach vorne in die Heizposition geschwenkt Dieser Vorgang kann je nach Materialstärke mehrfach durchgeführt werden 16 Service Wartung Grundsätzlich ist d...

Страница 6: ...yclable materials In accordance with the 2012 19 EU Waste Electrical and Electronic Equipment Direc tive SCHEU DENTAL will take back the unit free of charge if it is returned by free delivery Alternatively the machine can be sent to a designated collection point for recycling of electrical and electronic equipment 2 General Commissioning For optimal use of the device ensure to understand and follo...

Страница 7: ...e main socket protecting the electrical parts from overvoltage In order to exchange the fuses press the small clip underneath the power socket and remove the fuse box 8 For the device versions in 110 V 115 V these fuses can be found in two separate holders R that can be unscrewed For the device version in 230 V the fuse TAA REF 3100 940 for the device version in 115 V the fuse T10A REF 3101 341 an...

Страница 8: ...ft shall be opened These time intervals can be used for heating and bonding the occlusal surfaces using the heating support REF 3452 Once you have entered the corresponding heating time the heater with the attached heating support is swivelled forwards in the heating position This operation can be car ried out several times depending on the material thickness 16 Service maintenance The BIOSTAR dev...

Страница 9: ...ou recyclables conformément à la directive 2012 19 UE relative aux déchets d équipements électriques et électroniques SCHEU DENTAL reprendra l appareil gratuitement sous réserve que le retour est payé par l expéditeur Il peut également être collecté dans un point de collecte pour le recyclage des déchets électriques et électroniques 2 Mise en service générale Il est important de lire et observer l...

Страница 10: ...ensions Si vous souhaitez les échanger appuyez sur le petit clip sous la prise au courant et enlevez le porte fusible 8 Pour les versions en 100 V 115 V les fusibles se trouvent dans deux supports séparés R que vous pouvez dévisser Pour les appareils en 230 V le fusible TA4 REF 3100 940 est utilisé pour les appareils en 115 V c est le fusible T10A REF 3101 941 et pour les appareils en 100 V c est ...

Страница 11: ...res Veillez à ce que la chambre de pression soit fermée et l arbre de serrage soit ouvert Ces temps sont utilisés pour le chauffage ou la liaison des sur faces occlusales en utilisant le support de chauffe REF 3452 Le temps de chauffage saisi pivotez le radiateur avec le support de chauffe accroché vers l avant dans la position de chauffe Ce procédé peut être réalisé plusieurs fois selon l épaisse...

Страница 12: ...to se ha fabricado con materiales reutilizables o reciclables Conforme a la directiva sobre residuos de aparatos eléctricos y electrónicos 2012 19 UE SCHEU DENTAL acepta la devolución voluntaria del aparato de forma gratuita Cabe también la posibilidad de entregar los aparatos eléctricos y electrónicos en un punto de recogida para su reciclaje 2 Puesta en funcionamiento general Es preciso leer y s...

Страница 13: ...el sistema electrónico frente a subidas de tensión En caso de que sea necesario cambiarlos pulse el pequeño cierre situado bajo la toma de alimentación y saque el compartimento del fusible 8 En las versiones de 100 V 115 V los fusibles se encuentran en 2 sujeciones separadas R que se pueden atornillar Para aparatos de 230 V utilice el fusible T4A 3100 940 para los de 115 V el modelo T10A 3101 941 ...

Страница 14: ... o férulas Estos intervalos permiten el calentamiento o la unión de superficies oclusales con ayuda del soporte de calentamiento REF 3452 Tras la programación del tiempo de calentamiento correspondiente se gira la lámpara hacia delante con el soporte de calentamiento fijado a ella hasta la posición de calentamiento Dependiendo del grosor del material es posible realizar este proceso varias veces 1...

Страница 15: ...12 19 UE sui rifiuti di apparecchiature elettriche ed elettro niche SCHEU DENTAL ritira senza alcun costo l apparecchio usato con modalità di restituzione gratuita In alternativa esso può essere conferito a un centro di raccolta per il riciclaggio di apparecchiature elettriche ed elettroniche 2 Istruzioni generali per la messa in esercizio Per l utilizzo ottimale di BIOSTAR è necessario leggere e ...

Страница 16: ... si trovano anche due fusibili che proteggono l elettronica dell apparecchio dalle sovratensioni Per l eventuale sostituzione premere la piccola clip sotto la presa di rete ed estrarre lo scomparto portafusibili 8 Nella versione dell apparecchio a 100 V 115 V i fusibili sono collocati in 2 supporti separati R facilmente svitabili Per gli apparecchi da 230 V utilizzare il fusibile T4A 3100 940 per ...

Страница 17: ...mmazione speciale per la realizzazione di posizionatori protettori orali o splint di intervalli di tempo di 60 120 180 240 secondi Questi tempi servono al riscaldamento o all unione di superfici occlusali con l ausilio del supporto per il riscaldamento COD 3452 Al termine della programmazione del tempo di riscaldamento corrispondente la lampada con supporto per il riscaldamento agganciato viene or...

Страница 18: ...mer ein wenig eingefettet ist Nur so werden ein leichtes Verriegeln und ein geringer Verschleiß der Dichtung gewährleistet Eine verschlissene Dichtung kann durch einfaches Herausziehen aus der Hülsen halterung beim BIOSTAR getauscht werden Vor dem Einsetzen einer neuen Dichtung sollten Sie auch die äußere Seitenfläche der Dichtung einfetten Bild 4 Das Messingdruckstück f mit Schraube e Bild 2 könn...

Страница 19: ...schmutzung nicht mehr gängig sein Pflegehinweis Bei der Durchführung von nebenstehenden Pflegemaßnahmen am Tiefziehgerät achten Sie bitte darauf dass die Beweglichkeit des Ausgleichsrings jederzeit sichergestellt ist Der Ausgleichsring ist durch Eindrücken an einer Seite und gleichzeitiges Anheben auf der gegen überliegenden Seite aus der Druckkammer herauszu nehmen Auch die 4 Nieten mit den Druck...

Страница 20: ...nsation ring is not straight or not running well due to debris Maintenance hint Ensure the compensation ring is freely movable at any time when carrying out the maintenance advices given on the right Pull out balancing ring of pressure chamber by pressing on one side and lifting up on the opposite side simultaneously Remove the 4 rivets and pressure springs Clean and grease them using vaseline and...

Страница 21: ...odèle L anneau de compensation peut être incliné ou dur à cause des impuretés Astuce de maintenance Veiller à ce que la mobilité de l anneau de compen sation soit assurée à tout moment en effectuant les mesures de maintenance mentionnées ci contre Retirer l anneau l anneau de compensation de la chambre de pression en appuyant d un côté et en soulevant simultanément de l autre côté Retirer maintena...

Страница 22: ... de suciedad y está bloqueado Instrucción de manutención Fijarse en la movilidad del anillo de compensación en todo momento durante la ejecución de las medidas de manutención contiguas Sacar el anillo de compensación de la cámara de presión calcando un lado y levantando el de enfrente Ahora también se pueden sacar los 4 remaches con los muelles de compresión Limpiar y engrasar con vaselina las sup...

Страница 23: ...compensatore è inclinato o bloccato causato da insudicia mento Consiglio per la manutenzione Effettuando di tanto in tanto manutenuzione descritta qui a fianco l anello compensatore manterrà sempre la sua mobilità Premendo l anello compen satore da un lato esso si solleva all altro ed è facilmente estraibile dalla muffola Anche i quattro ribattini con le molle a compressione devono essere estratti...

Страница 24: ... une manipulation inappropriés Les cas de non respect des consignes de manipulation nettoyage et maintien Les cas de branchement inapproprié ou d installation incorrecte Les cas de force majeure ou d autres influences externes Les cas de modifications effectuées sur l appareil Les cas d utilisation des pièces détachées d un autre fabriquant Les cas de signes d usures dus à l âge n ayant pas d effe...

Страница 25: ...t dienen vorbehalten Nachdruck auch auszugsweise nur mit unserer schriftlichen Genehmigung Technical modification which serve for improvement are subject to our agreement Reproduction even in extracts is only allowed with our written approval ...

Страница 26: ...t été testées et approuvées sur le plan de la biologie cellulaire conformément aux normes internationales ISO 10993 et ISO 7405 E Nuestras planchas estan controladas y probadas citobiológicamente según las normas internacionales DIN EN ISO 10993 y ISO 7405 I I nostri dischi sono testati e approvati biologicamente secondo le norme internazioniali ISO 10993 y ISO 7405 S C H E U G R O U P SCHEU DENTA...

Отзывы: