![SCHEU 3001 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/scheu/3001/3001_operation-manual_1211761020.webp)
Measures to solve technical problems
1
2
3
Error
Probable reason
Solution
The foils are partially burned or too soft
You entered a wrong code or hea
ting-time for the foil.
Ensure to obser
ve the ba
tch specifica
tions given in the product
labelling.
Pressure moulding results are not sa
tisfying
You entered a wrong code or hea
ting-time for the foil.
Ensure to obser
ve the ba
tch specifica
tions given in the product
labelling.
No pressure build-up
When embedding the model in pellets,
the model cup has not
been filled up to the c
losing area.
The ma
terial rim is pulled out of the c
lamping ring during
thermoforming.
Please make sure tha
t only foils with a diameter of 125 mm are
used!
Fill the model cup with
pellets up to the c
losing
area.
Compensa
tion ring is not straight or not running well due to
debris.
Maintenance hint:
Ensure the compensa
tion ring is freely movable a
t an
y time
when carr
ying out the maintenance advices given on the right.
Pull out balancing ring of
pressure chamber by
pressing on one side and
lifting up on the opposite
side simultaneously
.
Remove the 4 rivets and
pressure springs.
Clean and grease them
using vaseline and refit
the ring.
Difficulties in c
losing pressure chamber
Pellets on the c
losing area of housing/model cup rim.
Before c
losing the pressure chamber make sure tha
t the c
lo
-
sing area is c
lean!
Difficulties in opening pressure chamber
The different steps for opening the pressure chamber ha
ve not
been carried out correctly
.
After depressurisa
tion and
5 seconds waiting time,
carr
y out the follo
wing
opera
tions one after
the other:
1
Turn the locking handle
180° backwards
2
Turn the bayonett c
losing ring to the left
3
Open the pressure chamber