Schambeck SFD GmbH
Refractive Index Detector RI2000
SFD101.01.3x.EN / 2014-07-10
6
http://www.schambeck-sfd.com
In case of the use of dangerous solvents pay attention to safety instructions
regarding this solvent.
Solvents should be degassed before use with a refractive index detector.
After use of salt containing solvents (such as buffers) the detector should be purged
with distilled water.
Make sure that the 6 bar (0.6 MPa) pressure limit of the flow cell is not exceeded.
Make sure that the 2 bar (0.2 MPa) pressure limit of the valve is not exceeded.
During operation the housing of the detector should be closed.
Do not use the detector in ambience of aggressive gases, very high humidity, strong
vibrations and strong changes in the ambience’s temperature.