Schambeck SFD GmbH
Refractive Index Detector RI2000
SFD101.01.3x.EN / 2014-07-10
43
http://www.schambeck-sfd.com
At this point flush (carefully!) the reference chamber and the sample chamber with at least
5 mL distilled water. It is recommended to use a disposable syringe during the calibration
to press the solutions into the INLET port. When the first step is finished, press the
ENTER
button. The display will change to the following view:
The signal is recorded and checked. In the display line two the difference voltage (Diff, -
0100 in this example) and the sum voltage (Summ, 5030 in this example) is shown.
If the signal is stable the display view will change to:
The valve is switched automatically into the measuring position. Now flush (carefully!) the
sample chamber of the flow cell at least three times (3 x 2ml) with the
test standard
solution
of exactly
343 mg glucose in 100 mL distilled water
before pressing the
ENTER
button to proceed with step two of the calibration process.