Schambeck SFD GmbH
Refractive Index Detector RI2000
SFD101.01.3x.EN / 2014-07-10
44
http://www.schambeck-sfd.com
After pressing the
ENTER
button the detector registers the current signal while the display
will show the different voltages as described before.
If the signal is stable the calibration factor is calculated and saved. The success of the
calibration process is confirmed by a displayed message.
Press the
ENTER
button to finish the calibration process. The new calibration factor is
shown in the upper right corner of the display. The four measured and calculated voltages
are shown in the display according to the ViewFine mode (
see chapter 15.2
). The calibration
is now finished.
In case of a problem during the calibration, an error message will be displayed.
In this case the RI2012 uses the old calibration factor for further measurements. Repeat
the process of calibration as describe before.