Schambeck SFD GmbH
Refractive Index Detector RI2000
SFD101.01.3x.EN / 2014-07-10
61
http://www.schambeck-sfd.com
The voltages are calculated according to the following formulas
Diff = Smpl
– Rfrn
Summ = Smpl + Rfrn
Use a multimeter to check the lamp voltage at the test point on the circuit board. For
this you need the serial number of your instrument (
see chapter 15.1
) to locate the
test points. If you serial number is 0312xxx the points are labelled “Lampe” and
“GND”. In case of a serial number of 0401xxx the points are labelled “LampeULmp”
and “GND”. In both cases the lamp voltage should be set 3.3
0.3 V. On pages 68
you can find a schematic view of the main circuit board with the marked
positions of the test points.
If the sum voltage is only a few millivolt the light bulb does not work. Check the power
cables of the light source to make sure that they are connected properly. If needed
check the light source (
see chapter 16.4
The sum voltage should be in the region of 5000
500 mV. If this is not the case
adjust the lamp voltage by turning the potentiometer R19 until the sum voltage is in
the region of 5000
500 mV.
Due to the adjustment of the light intensity it is necessary to check the lamp voltage
again. If this voltage is not in the region von 3.3
0.3 V the light source has to be re-
adjusted.
The optical system is adjusted properly if the sum voltage is 5000
500 mV and the
lamp voltage 3.3
0.3 V.