Schambeck SFD GmbH
Refractive Index Detector RI2000
SFD101.01.3x.EN / 2014-07-10
29
http://www.schambeck-sfd.com
14.5 S
ETTING
S
IGNAL
S
MOOTHING
To adjust the grade of smoothing of the detector signal press the menu button until the
following status line is shown in the display.
There are four different modes for signal smoothing:
RAW
No signal smoothing
FAST
Signal smoothing by calculating the average in a time interval of 0.4
seconds
MEDIUM
Signal smoothing by calculating the average in a time interval of 0.8
seconds
SLOW
Signal smoothing by calculating the average on a time interval of 1.6
seconds
The
ArrowUp
button (
▲
) changes to the next faster way of signal smoothing while the
ArrowDown button (
▼
) switches to the next slower mode. To accept the new settings press
the
ENTER
button.