Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
Calibration
15/01/14
10:00
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
Calibration
Probe test
9.6.2
Configuration
→
Diagnostic
→
Gas probe
KEY
FUNCTION
Enables the displayed interactive operations.
Also activates the interactive function visible on the left side of the screen.
Return to the previous page.
CONTEXT KEY
FUNCTION
Starts the test to check the seal on the gas probe.
Repeats the leak test for the gas probe.
Carrying out the leak test for the probe:
OK
→
→
Results:
Tightness
: the system is OK
Error:
Check that the probe is connected to input P-, check the seals of the pneumatic connectors and/or the seal of the water trap
and check that the test case is correctly positioned on the probe plunger.
CAUTION: a damaged probe tip may impair the test.
15/01/14
10:00
Diagnostics
Gas probe
Close the flue gas probe
Press OK to start
OK
Connect the gas probe equipped with the filter unit to the Instrument. Fully insert the probe
tip into the black cap provided with the Instrument as shown in the figure below:
15/01/14
10:00
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
OK
15/01/14
10:00
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
Calibration
Probe test
Result: leak
Black rubber cap
(provided)