33
9.2
Configuration
→
Instrument
PARAMETER
DESCRIPTION
In this sub-menu you can activate or deactivate the wireless communication of the instrument with a PC, smartphone
or tablet via Bluetooth
®
.
IF THE INSTRUMENT WORKS WITH BLUETOOTH
®
ACTIVATED, THE BATTERY LIFE IS
REDUCED TO 10 HOURS.
Time/Date
Permits the updating of the time and date. The format can be changed by selecting the EU (European) / USA
(American) mode.
Brightness
With the arrow keys you can increase or decrease the contrast of the display. This operation is also possible on the
first page at startup.
Pump
In this sub-menu you can switch the gas suction pump on or off. If the pump is on, it is also possible to read its
range in litres/minute. It is not possible to switch off the pump if the autozero cycle is in progress.
The CO sensor is protected by a pump that, if necessary, injects clean air to reduce the gas concentration on the
sensor. The function can be activated if a programmable threshold is exceeded or independently of the
concentration read by the instrument if one is aware of working with high levels of carbon monoxide.
DILUTION ONLY PLAYS A PROTECTIVE ROLE BECAUSE IT GREATLY DEGRADES THE
ACCURACY AND RESOLUTION OF THE MEASUREMENT.
Micromanometer
Allows you to configure the input of the micromanometer as a P+ or P- port. If you choose P- the pressure sign
will be inverted.
KEY
FUNCTION
Enables the displayed interactive operations.
Return to the previous page.
CONTEXT KEY
FUNCTION
◄
Select the available parameters.
OK
Enter in the selected parameter
►
Select the available parameters.
Time/Date
Brightness
Pump
Micromanometer
CO dilutor
CO dilutor