15
6
TECHNICAL FEATURES
6.1
Technical Features
Autozero:
Automatic autozero cycle with probe NOT inserted in the stack.
Dilution:
Increases the measurement range of the CO sensor to 100,000 ppm (10.00%).
Programmable to provide simple protection of the CO sensor by intervening when the threshold is
reached.
Default trip threshold set at 1500 ppm
Gas measuring sensors:
Up to 6 electrochemical sensors.
Auto diagnostics:
Verification of all features and internal sensors with anomaly reports.
Temperature measurement:
Double K thermocouple input with mini connector (ASTM E 1684-96) for differential temperature
measurement (send and return)
Ambient temperature measurement:
Through the internal sensor or through the T2 thermocouple input remote sensor
Fuel type:
12 factory pre-sets and 16 programmable by the user.
Power supply:
Li-Ion battery pack 3.7 V, 5200 mAh, with internal protection circuit
Battery charger:
External 5Vdc 2A battery charger with USB type A female con connection to the
instrument via the same serial communication cable supplied with the unit.
Charging time:
5 hours to recharge from 0% to 90% (6 hours for 100%). Recharging can also be done
by connecting the instrument to the computer; the instrument must be switched off and the
charging time varies depending on the current supplied by the computer, meaning the
process can last up to 12 hours.
Instrument life:
12 hours of continuous use.
Internal data memory:
2000 complete analyses with the possibility to enter date, time and name of client.
User data:
8 names of programmable operators.
Printed header:
Six lines of 24 characters, customisable by the user.
Screen:
Backlit 4.3" TFT colour display.
Communication port:
USB with TYPE B connector
Bluetooth
®
(depending on the model) :
Class 1 / Communication distance: <100 metres (in open field).
Line filter:
Includes replaceable cartridge, 99% efficiency for 20 um particles.
Extraction pump:
1.0 l/min draught in the chimney up to 135 hPa.
Range measurement:
Internal sensor to measure the range of the pump.
Water trap:
External to the instrument.
Carbon black:
When using an external, manual pump, the option to insert and print the opacity category
is available.
Efficiency (calculated):
0 – 100%
Condensing boiler efficiency:
Automatic recognition of condensing boiler, with calculation and printing of the
efficiency (0 - 120%) on P.C.I.
Environmental gases:
Measurement printing of ambient CO values.
Operating temperature:
-5°C to +45°C
Storage temperature:
-20°C to +50°C
Operating humidity range:
20% to 80% HR
Protection category:
IP42
Air pressure:
Atmospheric (from 800 to 1100 hPa)
External dimensions:
Analyser: 9.2 x 27.1 x 6.6 cm (L x H x D)
Rigid case: 50 x 39 x 13 cm (L x H x D)
Weight:
Analyser: approx. 0.9 kg
——————————————————————————————————————————————————
European standards:
Compliant with standards EN 50379-1 and EN 50379-2
European directives:
2014/30/UE CEM; 2014/35/EU Low voltage; 2011/65/UE RoHS II; 2012/19/EU WEEE