79
Each heater module can run a test inde-
pendent from the other heater modules
by recalling a program from the pro-
gram list and initiating a test. The pro-
gram list is common for all heater mod-
ules. When a test is initiated, the display
will show test screens from the single
module mode, such as entering an ID
number and bar graph prompt to add
the ideal sample size.
To test a sample in multi-mode:
§
Press the navigational
right
or
left
arrow
key to move the double bars to
the desired heater module.
§
Check current program. To change
program, press
Recall
and select a new
program if desired.
§
Press the
Start|Stop
key to initiate a
test. (the screens and functionality are
the same as in the single module mode).
§
After the initial weight is captured, the
display will return to the multi-mode
screen. The status line will show the
elapsed time of the test to indicate a
test is in progress.
§
At the end of the test, the status line
will display “Test Done” (or Pass or Fail
if target limits have been entered in this
program). The printer (optional) will
also print the result at the end of the
test.
Note
Certain functions are not available in
the multi-module mode including:
Interval print and graphing. However,
in the multi-module mode, one can test
one or more samples and still have
access to service for another heater
module which is not testing a sample.
To perform service options on any
heater module such as:
– Balance calibration
– Temperature adjustment
§
From the multi-module screen, press
the navigational
right
or
left arrow
key
to move the double bars to activate
the desired heater module.
§
Press the
Service
soft key to display the
Service menu.
§
Press the number key of the desired
option and press the
Enter
key.
The display will show these screens
as in the single module mode.