![Sartorius LMA110S Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/sartorius/lma110s/lma110s_operating-instructions-manual_1199621012.webp)
Assembly
The analyzer does not come fully
assembled. The control and heater
modules are placed next to each other
and may be placed on either side of
each other. Assemble the modules and
individual components in the following
order:
1. Open the heater hood using the handle
and install the pan shield so that the
center hole fits over the center ring of
the cavity base plate.
2. Next, place the stem of the pan support
through the center ring of the base
plate and into the hole of pan receiver.
3. Remove the foam from the heater
chamber.
4. Connect the power module cable
between the control module and the
heater module, observing the pin orien-
tation.
5. Connect the inter-module cable
between the control module (MOD
COMM 1 connector) and the heater
module, observing the connector orien-
tation.
6. If multiple heater modules are used,
install the additional inter-module
cables from each heater module to the
control module inter-module cable
receptacle (MOD COMM 1-2). Refer to
the chapter in this manual on multi-
mode operation for instructions specific
to operating multiple heater modules.
12