Data Output
LED Status Lights
The Mark 3 LTE heater module has two
LED lights on the front left corner that
visually indicate the status of the ana-
lyzer. The left LED shows the status of
the balance. The right LED indicates the
status of the heating function.
Each LED may have three colors:
red, green or amber. Red is generally
associated with a warning state and
shows that corrective action must be
taken. Green is generally associated with
a positive condition, indicating you can
proceed. Yellow generally indicates
a transitional period or that the test
is in progress.
20
Warmup
Balance LED
Heater LED
Warming up
Rapidly flashing RED
Rapidly flashing RED
Ready to test
Steady GREEN
Steady GREEN
Heater in conserve mode
Steady GREEN
Steady YELLOW
Standby
Balance LED
Heater LED
Sample in chamber
heating cavity (chamber)
Flashing YELLOW
Flashing GREEN
Test mode
Balance LED
Heater LED
Test in progress
Steady YELLOW
Steady YELLOW
Test done
Flashing YELLOW
Flashing GREEN
Sample loading
Below target
Flashing GREEN
Steady GREEN
Within target
Flashing YELLOW
Steady GREEN
Above target
Flashing RED
Steady GREEN
Diagnostics
Balance LED
Heater LED
Failed diagnostics
Flashing RED
Flashing RED