Choosing a Suitable Location
In order to ensure that the analyzer
functions properly, select a location that
meets the following requirements:
§
Permissible ambient environment
– Temperature:
+15 to +40°C (59°F to 104°F)
– Relative humidity:
10 – 85%, non-condensing
– Altitude:
0 to 2,000 meters (0 to ~ 6562 ft)
§
The analyzer is intended for indoor use.
§
Place the analyzer on a rigid, horizontal
surface, preferably free of vibration.
§
Avoid drafts and excessive temperature
fluctuation.
§
Locate the analyzer away from
flammable materials.
Connecting the Moisture Analyzer
to AC Power
This Class I Equipment (grounded
type) analyzer contains a universal
self-adjusting power supply rated to
100-120/200-250V~5/2.5A 50/60 Hz.
Check that this voltage matches your
local line voltage. If your voltage is not
within this rating, do not connect the
power cord to the power supply.
Danger
!
The analyzer may only be operated
using the original power cord supplied.
Please do not use it with an extension
cord. The wall socket must be grounded.
The AC Power mains supply fluctuations
are not to /–10% of the nomi-
nal supply voltage. The (Over Voltage)
Category II for transient over voltages
is EN 61010-1. The pollution Degree 2
is supplied EN 61010-1. The analyzer
is rated to Normal Protection (not
protected against harmful ingress of
moisture).
§
Insert the power cord that with the
analyzer into the power input receptacle
on the back of the heater module.
§
Then plug the power cord into a wall
socket.
§
Turn the analyzer on by moving the
power switch on the back of the analyzer
to the On position. The analyzer will
briefly proceed through a self-diagnostic
routine and then display the standby
screen. It will automatically enter a
warm-up period as noted on the screen.
Should the analyzer display any warn-
ings, consult the troubleshooting sec-
tion of this manual, or call Sartorius.
13