Modify Operators

our analyzer may be programmed with a custom list
of operators specific for your company. 
Operators can be alphanumeric and contain various
characters.

To change or set operator names:

§

Press the 

3

key from the Setup menu to display the

current list of operators.

§

Use the arrow soft keys, then press 

Enter

for any

operator slot to display the alphanumeric entry
screen. You may also press the 

Edit

soft key for the

highlighted name.

§

Enter the desired name, then press the 

Enter

key to

return to the Set Users/Operators Menu.

§

Line 20 is a blank line that cannot be edited.

§

When finished, press the 

Arrow

soft key to return to

the Setup menu.

Note

To enable the use of Operators and to have the
operator name on the result printout, Operator must
first be activated in the Results Printout Format
menu. When Operator is activated, then the 

User

soft key will be available from the Standby menu to
select an operator from the list of operators.

Note

Operator names will be deleted when downloading 
a new version of software.

46

   

 

 

•     

O p e r a t o r   s e t u p  

Edit 

Clear 

• 

0 5 / 0 1 / 2 0 1 1

  0 2 : 2 4   P M 

Содержание LMA110S

Страница 1: ...Operating Instructions Sartorius LMA110S Moisture Analyzer Model Mark 3 LTE Electronic Moisture Analyzer 98040 699 09 Rev D...

Страница 2: ...30 Units 31 Factors 32 Temperature 1 and 2 33 Time 1 and 2 34 Slope 34 Target Limits 39 Test Mode 40 Program Name 41 Create a New Program 42 Delete Program 42 Sort Programs 42 Configuring the Printout...

Страница 3: ...ting with infrared quartz tubes Precision electronic balance for accurate weighing and continuous weight moni toring during analysis Quarter VGA screen for easy viewing Large storage of methods for ea...

Страница 4: ...only for performing moisture analysis of samples Any improper use of the analyzer can endanger persons and may result in damage to the ana lyzer or other material assets Samples or containers e g beak...

Страница 5: ...in a secure place to ensure that it cannot be used for the time being Clean your moisture analyzer according to the cleaning instruc tions only see Cleaning Do not open the analyzer housing If the se...

Страница 6: ...arises in connec tion with this moisture analyzer Poisoning Caustic burns Substances containing toxic or caustic or corrosive substances These may be dried only under a fume hood The value for the lo...

Страница 7: ...he analyzer away from flammable materials If liquid is spilled into the analyzer unplug the Heating Module from the electrical supply immediately Analyzer is suitable for continuous operation Safety P...

Страница 8: ...trol module On initial power up the control module will automatically detect how many heater modules have been connected and are powered on If more than one heater module is detected the analyzer will...

Страница 9: ...hat you have received the following accessories Instructions for Recycling the Packaging To ensure adequate protection for safe shipment your moisture analyzer has been packaged to the extent necessar...

Страница 10: ...e cable receptacle 11 Main power switch 12 Main cord receptacle 13 Fuse 1 250 V 5A T 14 Fuse 2 250 V 5A T 15 Module power cable receptacle No Designation Control Module 16 Storage Bin 17 VGA display 1...

Страница 11: ...Control Module Connector Panel Heater Module Connector Panel Cables Connected to Control and Heater Modules 11 23 20 21 11 15 12 21 20 25 26 27 24 22 10 13 14 15 10...

Страница 12: ...the base plate and into the hole of pan receiver 3 Remove the foam from the heater chamber 4 Connect the power module cable between the control module and the heater module observing the pin orien tat...

Страница 13: ...d to the power supply Danger The analyzer may only be operated using the original power cord supplied Please do not use it with an extension cord The wall socket must be grounded The AC Power mains su...

Страница 14: ...o level Adjust control module so that it does not rock Note The heater module must be re leveled each time the analyzer is moved LMA110S Keypad Warm up To enable the analyzer to function optimally plu...

Страница 15: ...the key functions Dedicated keys Each key has a particular function Pressing the key once will initiate the desired action Initiates the testing of a sample Stops a test in progress u Press to finali...

Страница 16: ...les a highlighted selection between on and off Used to select a drying program from the Program menu Access to all setup parameters from the Setup menu Setup Recall On off Print Used to select an oper...

Страница 17: ...l display the program name date and time temperature and the last result if a test has been run Test This mode is entered when the Start Stop key is pressed to initiate a test The screen is similar in...

Страница 18: ...ected to perform statistical calculations 18 T e m p e r a t u r e t i m e D R Y M I L K P O W D E R 0 5 0 1 2 0 1 1 0 2 1 4 P M Print Result 210 30 0 5 1 0 1 5 2 0 2 5 3 0 3 5 4 0 4 5 5 0 min 25 67 V...

Страница 19: ...l return you to the Standby screen To return to the Standby screen press the bent arrow if one is available from that screen To choose a specific selection option or to enter a value A selection can b...

Страница 20: ...w generally indicates a transitional period or that the test is in progress 20 Warmup Balance LED Heater LED Warming up Rapidly flashing RED Rapidly flashing RED Ready to test Steady GREEN Steady GREE...

Страница 21: ...you may do this yourself by following the instruc tions under Setup in this manual Standby Screen During warm up your analyzer will enter the Stand by mode and the Standby screen will be displayed It...

Страница 22: ...s Weigh soft key to use as a balance Standby Temperature with a New Program When creating a new program the standby temperature setting from the previous drying program is used automatically by the an...

Страница 23: ...ure endpoint unit of measure standby temperature etc You will notice if you are in the default program or a specific material program by the program name at the top left of the display 23 Program name...

Страница 24: ...the Enter key for the number that precedes the program name As an alternative use the arrow keys to highlight the desired program name then press Enter The display will change to the Standby screen wi...

Страница 25: ...tart Stop key The display will change to the test mode as shown on the left The balance will automatically tare and show zero weight on the screen Note If you forgot to place the pan onto the balance...

Страница 26: ...e units as speci fied in the drying program Note To stop the test in progress press the Start Stop key The optional printer will begin to print as the test begins Printout information can be selected...

Страница 27: ...d Caution The sample is hot during and after the test Remove the sample with care by using the tweezers provided with the analyzer Do not touch any internal part of the heating chamber or pan shield a...

Страница 28: ...rogram Manage Menu The Program Manage Menu provides access to all setup functions associated with creating editing or managing sample drying programs To enter the Program Manage Menu From the Setup me...

Страница 29: ...the Standby screen the warning will show that the changes have not been saved To save an edited program Make changes then press the Save soft key to return to the Standby screen 29 D e f a u l t P r o...

Страница 30: ...99 Weight variable of endpoint criteria Window of time 0 1 99 9 minutes Time variable of endpoint criteria Calculation Actual slope Slope criteria met for endpoint Calc 1 Calculated endpoint Calc 2 C...

Страница 31: ...rrent unit will be shown Press a number key for the new unit of choice Note that there is an advanced option indicated by Factor on the soft key To change this feature see Factors When all the paramet...

Страница 32: ...e desired value and press the Enter key to return to the Factors screen Press the Arrow soft key to return to the Units screen result scale offset 32 C u r r e n t u n i t s M o i s t u r e 1 M o i s...

Страница 33: ...e Temp 1 or Temp 2 screen The current temperature will be shown Using the numeric keys type in the desired tempera ture and press Enter Note that there are some advanced options indicated by Ramp and...

Страница 34: ...keys type in the desired temperature then press the Enter key The display will return to the Tempera ture 1 or 2 screen Slope The slope function provides an auto matic endpoint for the test The slope...

Страница 35: ...l return to the Slope screen Endpoint Calculation There are three additional advanced slope settings called Calc 1 Calc 2 and Calc 3 which are algo rithms that calculate the endpoint prior to achievin...

Страница 36: ...soft key to return to the Program Menu Standby Temperature The standby temperature is the temperature that the analyzer chamber will be controlled to at the end of a test or prior to a test when a new...

Страница 37: ...he desired ideal weight then press the Enter key Note that there are some advanced options indicat ed by Lock and on the soft keys To change these features see Ideal Weight Percent Limit and Ideal Wei...

Страница 38: ...he Ideal weight percent screen Use the numeric keys to type in the desired time press the Enter key The display will return to the Ideal Weight screen Ideal Weight Lock When activated the ideal weight...

Страница 39: ...ture on and to enter limits Press the 8 key from the Program Menu to display the Target Limits screen The current target limits are shown Press the On Off soft key to toggle target limits On Press the...

Страница 40: ...solid samples The sample solids are first concentrated onto a filter pad placed in a vacuum apparatus During the test mode the operator will then be prompted to enter the number of milliliters of samp...

Страница 41: ...through the alphabet starting with A forward followed by symbols Pressing the down arrow scrolls through the symbols followed by letters in backwards order To change to lower case letters press the Lo...

Страница 42: ...to store the new program in the first available blank program slot Note Programs will be saved when down loading a new software version Delete Program At times you may want to delete a drying program...

Страница 43: ...Program Manage menu press the 6 key The current program will be printed Print Program List To print the entire list of programs From the Program Manage menu press the 7 key The list of programs will...

Страница 44: ...data to enter the data log From the Data Log menu press the On Off soft key to toggle data log to On To prevent data from entering the data log From the Data Log menu press the On Off soft key to togg...

Страница 45: ...key to view statistics of the selected results The display will show statistics including the average standard devia tion relative standard deviation the highest value and the lowest value To print t...

Страница 46: ...ter the desired name then press the Enter key to return to the Set Users Operators Menu Line 20 is a blank line that cannot be edited When finished press the Arrow soft key to return to the Setup menu...

Страница 47: ...graph during a test From the Percent Time graph press the Weight soft key To view the Temperature Time graph during a test From either the Percent Time or Weight Time graphs press the Temp soft key To...

Страница 48: ...s u l t s P r i n t o u t M e n u On Off 0 5 0 1 2 0 1 1 0 2 4 4 P M 1 H e a d e r 2 A n a l z y e r I n f o 3 O p e r a t o r 4 S a m p l e n u m b e r 5 C a l i b r a t i o n I n f o 6 P r o g r a...

Страница 49: ...r line 3 LMA110S Moisture Analyzer Module 1 M30604026 Operator John Smith Sample ID 1 Calibration external 06 01 2011 09 00 AM Last temp cal 05 13 2011 08 20 AM Program 1 Titon X UNITS Moisture FACTOR...

Страница 50: ...rn to the Results Printout menu screen If Numeric Increment is selected you will be prompted to enter a starting number before return ing to the Results Printout menu screen To set a print interval Fr...

Страница 51: ...r Set time At the Time screen press the Format soft key to select the 12 or 24 hour clock Use the numeric keys to type in the local time If you want to use the 12 hour clock press the AM PM soft key t...

Страница 52: ...Port menu screen Press the 3 key for Set up serial port to make final selections 8 no parity 1 stop bit 7 even parity 1 stop bit 7 odd parity 1 stop bit Finally press the 4 key to toggle CTS On or Of...

Страница 53: ...xx Float 0 1 99 99 SLOPETIME x xx 0 9 99 minutes 0 OFF SET CALC ACTUAL ACTUAL CALC1 CALC2 CALC3 CALC ACTUAL CALC1 CALC CALC1 CALC2 CALC CALC2 CALC3 CALC CALC3 SET DELAY x Integer 0 9 DELAY x 0 9 sec S...

Страница 54: ...IT ON GET MODE STANDARD SYRINGE CONCENTRATION MODE STANDARD GET PROGRAMNAME String PROGRAMNAME string System Commands SET SN xxxxxxxx Set controller SN SN M3123005A 9 characters SET MN xxxxxxxx Set mo...

Страница 55: ...BCD BCD mm dd yyyy hh mm AM PM GET LCT LCT EXTERNAL GET SLEEP SLEEP OFF GET SLEEPTIME SLEEPTIME 00 00 AM GET WAKE WAKE OFF GET WAKETIME WAKETIME 00 00 AM GET SLEEPDAY 1 7 SLEEPDAY 1 7 ON OFF GET WAKE...

Страница 56: ...A N G O R P T E S F F O E M A N G O R P F F O N O O F N I G O R P N O O F N I G O R P T E S F F O O F N I G O R P F F O N O S T H G I E W N O S T H G I E W T E S F F O S T H G I E W F F O N O E R U T...

Страница 57: ...DO ARCHIVE LOAD ARCHIVE LOAD DO TARE TARING DO CLROPERATORS CLROPERATORS DO CLRDATALOG CLRDATALOG DO KEY 0 KEY 0 DO KEY 1 KEY 1 DO KEY 2 KEY 2 DO KEY 3 KEY 3 DO KEY 4 KEY 4 DO KEY 5 KEY 5 DO KEY 6 KEY...

Страница 58: ...Weight 6 mg L 4 Slope Time X X minutes 5 Slope Percentage X XXX 6 Slope Calculation 0 Actual 1 Calc 1 2 Calc 2 3 Calc 3 7 Standby Temperature 165C max 8 Equilibration Time X minutes 9 Ideal Weight XX...

Страница 59: ...05 0 0 0 60 0 10 0 050 65 FACTORY 1 2 M 30 0 110 0 120 0 110 0 20 0 010 65 FACTORY 2 3 S 2 0 135 0 0 0 130 0 10 0 100 65 FACTORY 3 4 W 5 0 105 130 300 300 60 0 0 0 000 65 FACTORY 4 5 L 2 0 135 0 0 0 1...

Страница 60: ...ain the standby temperature while the analyzer is not being used enter 0 minutes To set or change the sleep and wake mode From the Conserve Mode screen press 2 or 3 key to show the Sleep mode or Wake...

Страница 61: ...ghlight Security mode using the navigation keys and press Enter to display the Security menu Activate and Deactivate Security To activate security From the Security menu screen the press 1 for Set Sec...

Страница 62: ...Press the respective number key to toggle On or Off or highlight the selection and press the On Off soft key Press the Arrow soft key to return the Security menu Security must be set On in the Securit...

Страница 63: ...earing the data log causes a permanent loss of data Clear Operators At times you may want to clear all operators Operator 20 will always remain a blank line Security must be deactivated to perform thi...

Страница 64: ...well as clear all programs and any user setup parameters Manage Program Storage The analyzer has the capacity to store a custom set of programs in memory In the event that the user selectable program...

Страница 65: ...the Manage program storage screen Press the 2 key for Save user programs to archive and display the confirmation screen Note Saving user programs to archive will overwrite any existing programs that m...

Страница 66: ...Manage program storage and press the Enter key to dis play the Manage program storage screen Mark 2 only press Setup Setup 2 menu Prg Xmit and External input Press the 4 key for transfer programs Pre...

Страница 67: ...vigation keys and press Enter to display the Audio Video screen To change the key press audio beep From the Audio Video screen press the 1 key to toggle On or Off or highlight Key press beep and press...

Страница 68: ...ckground To change the video color scheme From the Audio Video screen press the 5 key for Video color scheme to select either black on white or white on black Setting the Language 6 languages are avai...

Страница 69: ...ddress To view the system screen From the Setup menu screen press the key or highlight Service using the navigation keys and press Enter to display the Service menu screen To view and print system inf...

Страница 70: ...calibrated manually with a known external calibration weight The analyzer also allows you to set a calibration interval that will remind you when calibration is due To access the balance calibration...

Страница 71: ...support and close the hood The analyzer will automatically recognize hood clo sure wait for stability and begin the calibration and adjustment procedure Note Permissible external calibration weights i...

Страница 72: ...ule When this interval expires the operator will be prompted on the screen to calibrate the analyzer However this message will not prevent testing a sample or any other operation of the analyzer Note...

Страница 73: ...se items if you are wearing protective gloves Otherwise leave the heater hood open and let these items cool down before attempting to remove them Place the temperature bowl fixture in the same place a...

Страница 74: ...ent process takes 60 minutes The Temp adjust done screen will show both the pre adjusted temperature along with the postad justed temperature for both calibration temperatures The time and date of las...

Страница 75: ...ure and the actual temperature that was measured by the temperature bowl fixture Press the Print soft key to print a temperature verification report optional If the agreement of the selected temperatu...

Страница 76: ...on or temperature adjustment should be performed To enter a reference fixture number From the Adjust Temperature screen press the 4 for Reference fixture The screen will change to have you enter the r...

Страница 77: ...Press the 2 key for Concentration The display will return to the Program menu To test samples in the Concentration mode Open the heater hood place the filter pad onto a pan and close the hood Press t...

Страница 78: ...Mod Comm ports have a heater module attached If more than one is connected the screen will display in the Multi module mode Note that a double set of bars shows the active heater module Multi module...

Страница 79: ...e screen The status line will show the elapsed time of the test to indicate a test is in progress At the end of the test the status line will display Test Done or Pass or Fail if target limits have be...

Страница 80: ...not immerse the analyzer in water or any other cleaning solution Regular cleaning of the analyzer with a mild detergent and a soft cloth should include the following Clean the outer surface of the an...

Страница 81: ...d waste or brought to collection centers run by local public disposal operations not even by small commercial operators In countries that are not members of the European Economic Area EEA or where no...

Страница 82: ...maximum programmable limits and screen alert when test done Program Storage 60 programs with unique settings alphanumeric naming Result Storage 999 results with statistics average SD RSD high and low...

Страница 83: ...CC01 MK5 Mark 3 to PC Ethernet cable YCC01 MK2 Extra long Inter Module cable 30 4th HM 69TC0131 Large Pan Support 69TC0289 Mark 3 Cal Val Kit YVK02MA 8 ml Pipette Box 400 YAT02MA 8 ml Pipette New Styl...

Страница 84: ...NT Electrical CLASS 8721 85 ELECTRICAL EQUIPMENT FOR LABORATORY USE Certified to US Standards Note The operator shall be responsible for any modifications to Sartorius equipment and for connections of...

Страница 85: ...85 Notes...

Страница 86: ...86 Notes...

Страница 87: ...87...

Страница 88: ...JW W3A000 CP KT Publication No 98040 699 09 Rev D Sartorius Corporation 5 Orville Dr Bohemia NY 11716 info sartorius omnimark com www sartorius us Tel 800 835 3211 Fax 631 253 5263...

Отзывы: