![Sartorius LMA110S Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/sartorius/lma110s/lma110s_operating-instructions-manual_1199621059.webp)
Mark 2 Program Download Format
Field
Description
1
Program Number (#XXX)
2
Units (M,S,W,V,L)
3
Ideal Weight (XX.X, Negative value indicates ideal wt. loading locked 90-110%)
4
Temp 1 (XXX)
5
Temp 2 (XXX)
6
Time 1 (Seconds)
7
Time 2 (Seconds)
8
Standby Temp (XXX)
9
Slope Mode (0=Actual, 1=Calc1, 2=Calc2, 3=Calc3)
10
Slope Time (Tenths of minutes, example: 50=5.0 minutes)
11
Slope % (XX.XXX)
12
Start Delay of Initial Wt. Capture*
13
Program Name (19 characters)
Output Example (Factory Defaults):
# 1,M, 5.0,105, 0, 0, 0, 60,0,10, 0.050,65,FACTORY 1
# 2,M,30.0,110, 0, 120, 0,110,0,20, 0.010,65,FACTORY 2
# 3,S, 2.0,135, 0, 0, 0,130,0,10, 0.100,65,FACTORY 3
# 4,W, 5.0,105,130, 300, 300, 60,0, 0, 0.000,65,FACTORY 4
# 5,L, 2.0,135, 0, 0, 0,100,0,10, 0.050,65,FACTORY 5
/r/n
* Field 12 - Start Delay of Initial Wt. Capture Bit 0...Sample ID 0=Off 1=On
Sample ID Off
Sample ID On
Time(s)
Value
Hex
Time (s)
Value
Hex
0
0
0000 0000
0
1
0000 0001
1
16
0001 0000
1
17
0001 0001
2
32
0010 0000
2
33
0010 0001
3
48
0011 0000
3
49
0011 0001
4
64
0100 0000
4
65
0100 0001
5
80
0101 0000
5
81
0101 0001
6
96
0110 0000
6
97
0110 0001
7
112
0111 0000
7
113
0111 0001
8
128
1000 0000
8
129
1000 0001
9
144
1001 0000
9
145
1001 0001
59