To initiate an automatic temperature calibration and
adjustment, press the
5
for Automatic Temp Cal.
The screen will change to the Temperature screen.
§
The analyzer will automatically begin at two
temperatures (80°C and 150°C). Each of Cal 1 and
Cal 2 temperature steps will take 20 minutes.
The time elapsed is shown on the screen. As soon as
the Cal 1 temperature phase has expired, Cal 2 will
begin automatically.
§
When the Cal 2 phase has elapsed, the screen
will change to show that adjustment has been
completed.
§
The analyzer will then verify the adjustment by
checking at 130°C for 20 minutes. The entire temp
calibration and adjustment process takes
60 minutes.
§
The “Temp adjust done” screen will show both the
pre-adjusted temperature, along with the postad-
justed temperature for both calibration
temperatures.
§
The time and date of last temperature calibration
and adjustment is automatically updated on System
menu.
§
Pressing the
Arrow
soft key anytime during the
temperature adjustment will abort the “auto temp
cal” process. The screen will return to the “Auto
temp cal” screen without adjusting the analyzer
temperature calibration.
A u t o t e m p c a l
Cal 1
Current
80
71
1 . 5 m i n u t e s
C a l B o w l T e m p : 7 5
T i m e = 2 0 m i n u t e s
0 5 / 0 1 / 2 0 1 1
0 2 : 4 0 P M
A u t o t e m p c a l
Standby
Current
60
124
C o e f f i c i e n t s s e t
C a l i b r a t i o n p a s s e d v e r i f i c a t i o n
C a l i b r a t i o n p a s s e d
0 5 / 0 1 / 2 0 1 1
0 2 : 4 4 P M