Temperature 1 and 2
The analyzer will dry the sample at one or two
temperatures set in one degree of Celsius incre-
ments. A second temperature is optional and is nor-
mally set to Off. See the Applications section for a
description of when to use a second temperature for
two-step drying.
To set or change Temperature 1 or Temperature 2:
§
Press the
2
or
3
key for Temp 1 or Temp 2,
respectively. The display will change to the Temp 1
or Temp 2 screen. The current temperature will be
shown.
§
Using the numeric keys, type in the desired tempera-
ture and press
Enter
.
§
Note that there are some advanced options indicated
by Ramp and Time 1 (or 2) on the soft keys. To
change these features see, “Time 1 and 2.”
§
When all the parameters for Temperature 1 or 2
have been changed, press the
Enter
again or the
Arrow
key to accept the changes. The display will
return to the Program Menu.
33
T e m p e r a t u r e : 1 0 5 C
M a x . T e m p e r a t u r e : 2 1 0 C
T i m e : 9 9 . 9 m i n u t e s
T e m p e r a t u r e 1
D e f a u l t
| C
Clear
Time
0 5 / 0 1 / 2 0 1 1 0 2 : 2 1 P M