![Sartorius CAIS1 Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/sartorius/cais1/cais1_operating-instructions-manual_1199616089.webp)
Net Total Formulation
R
(Combics 2)
With this application, you can weigh in different components up to a defi ned total.
Each component is saved in the net-total memory (
Applic.3
menu).
Features
– Weigh in up to 999 components in series
– Net total formulation
cannot
be combined with level 1 and 2 applications
(
Applic.1
,
Applic.2
).
– Current component number displayed in the number line (indicating the
component to be added)
– Toggle the display from “component mode” to “additive mode” by pressing the
w
key
–
Component mode
: Display the weight of the component currently on the
platform (for 1 second after it is saved; then the platform is tared)
–
Additive mode
: Display the weight of all components on the platform (after
it is saved, the net weight of the last component added is displayed briefl y)
– Toggle to a second weighing platform while weighing in
– Activate Info mode via the
I
key
– Automatic component printout when it is saved.
Setting:
Applic./Applic.3/NET TOT
menu item 3.17.
Printout
If the 3.17.2 menu item is selected, the entire component record is printed.
If the 3.17.3 menu item is selected, the following items are generated only once for
the fi rst component if they have been confi gured:
Blank line, dash line, date, time, ID1 through ID6, header lines 1 and 2. For
subsequent components, each “component” item (“Comp xx”) is followed by a
blank line.
– Automatic taring of container weight.
Setting:
Applic.
/
aut.tare
menu item 3.7
– Restore factory default settings
Setting:
Applic.
/
Reset
menu item 9.1
Preparation
t
Open the
APPLIC. / APPLIC.3 /NET TOT
menu.
Available parameter settings
* = Factory setting
MIN.INIT
Minimum load for initialization
3.6
1 digit
1 digit
3.6.1*
2 digit
2 digits
3.6.2
5 digit
5 digits
3.6.3
10 digt.
10 digits
3.6.4
20 digt.
20 digits
3.6.5
50 digt.
50 digits
3.6.6
100 dig.
100 digits
3.6.7
200 dig.
200 digits
3.6.8
500 dig.
500 digits
3.6.9
1000 dI.
1000 digits
3.6.10
PRT.SAV.
Individual/Component printout when saved
3.17.
OFF
Automatic printing off
3.17.1
EACH.TIM.
Print the entire standard print
confi guration every time the
O
key
is pressed 3.17.2*
ONCE
Print the entire standard print
confi guration once with the
O
key 3.17.3
Operating Instructions Combics Indicators
89
Operation
Содержание CAIS1
Страница 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity Declaration of Conformity...
Страница 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Страница 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Страница 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Страница 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Страница 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Страница 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Страница 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Страница 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Страница 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Страница 158: ...158 Operating Instructions Combics Indicators...
Страница 159: ...Operating Instructions Combics Indicators 159...