In the
Applic./Applic.3/CLASS./Param.2/OUTP.ACT
menu, menu item 4.7,
you can make the following settings for the control outputs:
– switched off
– always on
– on at stability
The “SET” output normally changes its voltage level when the current weight
exceeds the minimum load. Alternatively, you can assign the “Ready for use”
function to this port.
Setting:
Applic./Applic.3/CLASS./PARAM.2/CTRL.SET
menu item 4.3.
Example 2
: There should be three classes.
Configuration: The “Classification" application is selected, and printout has been
set up.
O
t
Enter the class limits using the
O
key.
.
1 1 0
t
Enter the upper limit for Class 1 using the keypad (in this example, 110 g)
.
+
O
t
Save the upper limit for Class 1.
1 3 0
t
Enter the upper limit for Class 1 using the keypad (in this example, 110 g).
O
t
Save the upper limit for Class 1.
t
Place the sample on the weighing platform.
Operating Instructions Combics Indicators
83
Operation
Содержание CAIS1
Страница 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity Declaration of Conformity...
Страница 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Страница 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Страница 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Страница 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Страница 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Страница 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Страница 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Страница 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Страница 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Страница 158: ...158 Operating Instructions Combics Indicators...
Страница 159: ...Operating Instructions Combics Indicators 159...