Calibration, Adjustment
Purpose
Perform
calibration
to determine the difference between the value displayed and
the actual weight on the platform. Calibration does not entail making any changes
within the weighing equipment.
During
adjustment
, the difference between the measured value displayed and the
true weight of a sample is corrected, or is reduced to an allowable level within
maximum permissible error limits.
Configuration for Use in Legal Metrology
Configuration of the weighing instrument for use in legal metrology is set by a
switch. The switch is located on the back of the weighing platform and covered by
a protective cap.
Using a verified scale in legal metrology in the EU:
The Type-Approval Certificate for verified scales is only valid for non-automatic
weighing instruments. For automatic operation with or without additional,
integrated equipment, please follow the applicable national regulations for the
installation location.
Externally connected IS scales: Before use in legal metrology, the scale should be calibrated via the internal
calibration equipment at the installation location: see the “Internal Calibration"
section in this chapter.
3
The temperature range (°C) listed on the ID label should not be exceeded during
operation.
For Servicing: External calibration for verified scales of accuracy class
l
–
External calibration is blocked in legal metrology (switch cover is sealed)
–
External calibration is only possible by removing the seal. If the seal is broken,
the validity of verification will become void and you must have your scale
re-verified.
Using a verified scale in legal metrology with internal calibration equipment:
t
Before use in legal metrology, the “internal calibration" function should be
carried out at the installation location:
Open the menu access switch
The menu access switch is located on the back of the indicator right next to the
weighing platform connection.
t
Remove the cap.
t
Slide the switch to the left (= “open" position, for use in legal metrology).
Operating Instructions Combics Indicators
47
Operation
Содержание CAIS1
Страница 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity Declaration of Conformity...
Страница 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Страница 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Страница 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Страница 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Страница 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Страница 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Страница 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Страница 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Страница 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Страница 158: ...158 Operating Instructions Combics Indicators...
Страница 159: ...Operating Instructions Combics Indicators 159...