With this application, you can calculate averages from several weighing operations.
It is used when either the object to be weighed (e.g. animals) or the environment
during weighing are unstable. Selection and settings in the Applic.
/ Applic.1 /
ANIM.WG
menu.
Features
– Averaging started manually or automatically (
.../start
menu item 3.18).
With manual start selected, the averaging routine begins when you press a key
(provided the start conditions are met).
With automatic start selected, the application begins when you place the first
load on the platform (provided the start conditions are met).
– Enter the number of sub-weighing operations using the keypad
– Use the
r
key to select the number of measurements for averaging
– Activate Info mode via the
I
key
– Toggle the display from “result of last measurement” to “current weight” by
pressing the
w
key
– Automatic results printout (
menu item 3.20).
– Automatic taring of container weight (
APPLIC./aut.tare
menu item 3.7).
– Automatic start of averaging when the scale is turned on and a sample placed
on the platform (provided start conditions are met,
APPLIC./aut.Strt
menu
item 3.8).
Exit application, delete parameters
The value of the reference sample weight in the reference memory remains active
until it is deleted via the
c
key or overwritten or until the application is changed.
The reference sample weight also remains saved after the scale is turned off.
You can assign different functions to the
c
key for deleting applications. When
you clear applications, you can delete either the data stored for all applications or
just selected data stored for the active application.
Setting:
Applic./CLER.CF
menu item 3.24
Tare function:
If you store a tare (weight value) by pressing the
)
key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting:
Applic. / TARE.FNC
menu item 3.25.1 (factory setting)
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting:
Applic. / TARE.FNC
menu item 3.25.2
Restore factory default settings:
Applic./Reset
menu item 9.1.
Number of measurements for average
You can enter the desired number of weight measurements to determine the average
using the keypad. This value remains active until it is overwritten by another value. It
also remains in memory when you switch to a different application program, or turn
off the scale.
Application start
There are three ways to start the averaging routine:
– Manual start with preset number of sub-weighing operations: Place the sample
on the platform and press the
O
key
– Manual start with user-defined number of sub-weighing operations: Place the
sample on the platform and enter the number of weighing operations using the
keypad. Press the
r
key to save the number entered and begin weighing.
– Automatic start with preset number of sub-weighing operations: Measurement
begins when you place the first sample on the platform, provided the start
conditions are met.
Preparation
t
Open the
APPLIC. / APPLIC.1 /ANIM.WG
menu.
Operating Instructions Combics Indicators
63
Operation
Содержание CAIS1
Страница 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity Declaration of Conformity...
Страница 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Страница 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Страница 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Страница 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Страница 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Страница 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Страница 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Страница 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Страница 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Страница 158: ...158 Operating Instructions Combics Indicators...
Страница 159: ...Operating Instructions Combics Indicators 159...