Adjustment Without Weights
In the Service menu, adjustment without weights can be carried out by entering the
characteristic data of the load cells.
3
Adjustment without weights may not be carried out on weighing equipment used
in legal metrology.
Setup information
– Adjustment without weights is only possible when the menu access switch is
open in the Service menu.
– When the Service mode is activated, you can enter the necessary parameters for
adjustment without weights in the
Setup
menu under “
WP-1
” for the first
weighing platform and
COm1 / WP-2
,
UNICOM / WP-2
or
COm-WP
for the second
weighing platform. The settings are made in the corresponding Setup menu
under menu item 1.19.
– The “Nominal load” parameter must be entered in the kg unit.
– The “Resolution" parameter must be entered in the kg unit and must correspond
to the scale interval d entered for the ADC configuration.
– The “Sensitivity” parameter is entered in mV/V (see the data sheet for the value).
h
The data entered are saved by selecting menu item 1.19.7. After saving, the data will
no longer be able to be read.
Procedure
t
Open menu access switch.
t
Activate the Service mode.
t
Select the weighing platform.
t
Enter the nominal load of the load cell(s) in kg under menu item 1.19.1. If the
weighing platform has multiple load cells, the nominal load must be multiplied
accordingly (e.g. 4 load cells, each of which has a capacity of 50 kg, will produce
a nominal load of 200 kg)
t
Enter the resolution in kg under menu item 1.19.2. The value must correspond
to the scale interval d entered under menu item 11.4.1.
t
Enter the sensitivity of the load cells in mV/V under menu item 1.19.3.
For weighing platforms with multiple load cells: Enter the individual values of
the load cells in 1.19.3 to 1.19.6 or enter the average of all load cells in 1.19.3.
t
Save the values for adjustment without weighing under menu item 1.19.7.
t
Close the menu access switch.
Menu structure for adjustment without weights
ADJ.W/O.W
Adjustment without weights (entering the characteristic data of the load cell(s))
1.19.
NOM.LOAD
Nominal load
1.19.1
RESOLUT
Resolution
1.19.2
SENSIT.1
Sensitivity in mV/V for cell 1 (or average value for all load cells)
1.19.3
SENSIT.2
Sensitivity in mV/V for cell 2
1.19.4
SENSIT.3
Sensitivity in mV/V for cell 3
1.19.5
SENSIT.4
Sensitivity in mV/V for cell 4
1.19.6
SAVE
Save values for 1. 19
1.19.7
30
Operating Instructions Combics Indicators
Configuring Weighing Platforms
Содержание CAIS1
Страница 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity Declaration of Conformity...
Страница 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Страница 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Страница 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Страница 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Страница 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Страница 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Страница 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Страница 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Страница 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Страница 158: ...158 Operating Instructions Combics Indicators...
Страница 159: ...Operating Instructions Combics Indicators 159...