![Sartorius CAIS1 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/sartorius/cais1/cais1_operating-instructions-manual_1199616017.webp)
Configuring Weighing Platforms
Service Mode
Purpose
The Service mode enables access to additional menu items in the Setup menu
(
setup
) which are not displayed when the Service mode is not active. The most
important calibration and adjustment work for the indicator and for the connected
weighing platform can be carried out in the Service menu, e.g. ADC configuration.
When the Service mode is active, an “
S
” is shown in the top right-hand corner of the
display. To deactivate the Service mode, restart the indicator (turn the indicator off
and back on again).
In Service mode, the
Setup
menu is expanded with the following parameters after
entering the user password:
–
S-DATE
for entering the next service date
–
SER.NO
for entering the device serial number
–
MODEL
with the
model description
–
S-SQMIN
–
ALIB.MEM
for deleting the
Alibi memory
The Setup menu for
WP1
and
WP2
can be extended to include the following setting
options:
param1
CAL./ADJ.
Calibration, adjustment
1.9.
Internal linearization (for WP-2 only)
1.9.5
CAL.EXT
External linearization with default weights
1.9.6
CAL.E.USR.
External linearization with user-defined weights (entered under 1.18)
1.9.7
SET.PREL.
Setting the preload
1.9.8
DEL.PREL.
Clearing the preload
1.9.9
HND.XT / CAL./ADJ.
Entering the adjustment and linearization weights
1.18.
LIN. WT.1
Entering the lin. weight 1
1.18.2
LIN. WT.2
Entering the lin. weight 2
1.18.3
LIN. WT.3
Entering the lin. weight 3
1.18.4
LIN. WT.4
Entering the lin. weight 4
1.18.5
ADJ.W/O.W
Adjustment without weights (entering the characteristic data of the load cell(s))
1.19.
NOM.LOAD
Nominal load
1.19.1
RESOLUT
Resolution
1.19.2
SENSIT.1
Sensitivity in mV/V for cell 1 (or average value for all load cells)
1.19.3
SENSIT.2
Sensitivity in mV/V for cell 2
1.19.4
SENSIT.3
Sensitivity in mV/V for cell 3
1.19.5
SENSIT.4
Sensitivity in mV/V for cell 4
1.19.6
SAVE
Save values for 1. 19
1.19.7
GEOG.DAT
Adjustment location (geographical data; or alternatively
1.20.
the gravitational acceleration at the place of installation)
LATITUD
Latitude in degrees
1 20.1
ALTITUD
Elevation in meters above sea level
1 20.2
GRAVITY
. Gravitational acceleration
1 20.3
SAVE
. Save values for 1. 20
1 20.4
ADC settings (menu see page 19)
11
Applying the serial number of the IS weighing platform
12.1
(verified weighing platform at WP2)
Apply the serial number
12.1.1
Inactive (standard WP)
12.1.2
Operating Instructions Combics Indicators
17
Configuring Weighing Platforms
Содержание CAIS1
Страница 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity Declaration of Conformity...
Страница 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Страница 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Страница 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Страница 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Страница 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Страница 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Страница 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Страница 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Страница 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Страница 158: ...158 Operating Instructions Combics Indicators...
Страница 159: ...Operating Instructions Combics Indicators 159...