It is used to set the offset value of the current measurement window. “Disp Offset” is only valid for the current
measurement window. When the “Disp Offset” is enabled, the prompt “Disp Offset On” will be displayed at the upper
right corner of the screen.
“Disp Offset" is only valid for digit display, and it doesn't work in the waveform display mode. All settings below the
[Measure] menu don
‟
t work in the waveform display mode.
[Chan Offset] below[Channel]is valid for the current channel. The effects of “Chan Offset” and “Disp Offset” can be
superimposed, e.g. when the current sensor is used to measure a 0.0 dBm signal: if “Chan Offset” is set at 10.0 dBm and
“Chan Offset” is in On state, the power result will be displayed as
10.0 dBm; if “Disp Offset” is set at 10.0 dBm and “Disp Offset” is in On state, the power result will be displayed as 20.0
dBm.
[State Off On]
It is used to enable or disable the display offset in the current measurement window.
[Offset]
It is used to set the display offset of the current measurement window.
To increase the measurement result , the “Offset” should be positive; otherwise it should be negative.
[Feed]
It is used to select which channel is involved in the result displayed in the current measurement window. At the same
time, it displays “Gate”, “Meas”, “Comb”, etc.
[Channel]
It is used to select the channel corresponding to the current window. Press [Channel] to switch between Channel A and
Channel B.
[Gate]
It is used to select the gate for the current measurement window. Press [Gate] to enter the next menu. There are four
optional gates in total. A prompt of the gate is displayed in the measurement window. For example, when Gate 1 of
Channel A is selected, “A1” will be displayed at the upper left corner of the digit display window.
At most four different gates are available. The specific settings area completed through [Channel]>[Channel A]>[Gates
Set].
The “Gate” is defined for waveform display, and can be used to set various power parameters of the waveform relative to
the trigger position within a certain period of time
[Meas]
It is used to select the result displayed in the current window as the measurement type corresponding to the current gate.
The menu includes five options, i.e. “AVG PWR”, “Peak PWR”, “PAR”, “Min PWR” and “Max/min”, with “AVG”,
“Peak”, “PAR”, “Min” and “Max/min” marked at the upper left corner of the current digit display window.
[AVG PWR]