CAUTION
Rotate the
screw
Keep
it
unmoved
Check the connectors so as not to damage the instrument ports
Any damaged connector may damage the good one connected to it even if it is connected for the
first time. To protect the ports of the signal generator, it is necessary to check the connectors before
operating them.
2)
Connection
The connectors should be checked and cleaned before measurement and connection to ensure
that they are clean and undamaged. It is necessary to wear an anti-static wrist strap during
connection. The correct connection methods and procedures are as follows:
Step 1. As shown in Fig.2.4, align the axis of both devices to be connected to ensure the
pins of male connector can concentrically slide into the holes of female connect;
Fig.2.4 Coaxial Alignment of Interconnected Connectors
Step 2. As shown in Fig.2.5, move both connectors straight together, so that they can be
connected smoothly; rotate the threaded sleeve of connector (rather than the connector itself)
until it is tightened; during connection, there can be no relative rotary motion both
connectors;
Fig.2.5 Connection
Step 3. As shown in Fig.2.6, use a torque wrench to complete the final connection. It should
be noted that the torque wrench should not exceed the starting kick point. An auxiliary
wrench can be used to prevent the connectors from rotation.